Sterility Duration of Preprimed Extracorporeal Membrane Oxygenation Circuits
Objectives:
There is a lack of standardization and supporting data regarding the duration preassembled and preprimed extracorporeal membrane oxygenation (ECMO) circuits are expected to be sterile. Therefore, the purpose of this study was to prospectively evaluate whether preassembled and preprimed ECMO circuits could maintain sterility for a period up to 65 days.
Design:
Four ECMO circuits (2 neonatal/pediatric¼" and 2 adolescent/adult ⅜ ") were assembled and primed under sterile conditions and maintained at room temperature. Culture samples were obtained from each circuit and plated within 1 hour. Culture samples were obtained on day 0 when assembled and primed then every 5 days up to day 65. Samples were plated on several different media including the following: blood agar plate: trypticase soy agar with 5% sheep blood, MacConkey agar, and thioglycollate broth then incubated at 35°C for 3 days.
Results:
All cultures obtained from the priming solution from of the¼" and ⅜ " ECMO circuits produced no microbial or fungal growth for the 65-day study period.
Conclusion:
These pilot data suggest preprimed ECMO circuits may maintain sterility for a period up to 65 days. Additional studies evaluating a larger number of ECMO circuits are needed to confirm these findings.
More Related Videos
03:40Point-of-Care Ultrasound for Peripheral Veno-Arterial Extracorporeal Membrane Oxygenation Without Left Ventricular Venting
Published on: January 17, 2025
13:56Modeling Biological Membranes with Circuit Boards and Measuring Electrical Signals in Axons: Student Laboratory Exercises
Published on: January 18, 2011
Related Concept Videos
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Extracorporeal Removal of Drugs: Hemoperfusion and Hemofiltration
Second-Order Circuits
Input signals typically originate from voltage or current sources, with the output often representing voltage across the capacitor and/or current through the inductor. For example, in...
First-Order Circuits
One common example of a first-order circuit is the RC (resistor-capacitor) circuit. These circuits are used in relaxation oscillators such as neon lamp oscillator circuits. When voltage is...