Reliability of virtual non-contrast computed tomography angiography: comparing it with the real deal
Leena Lehti1,2, Marcus Söderberg3, Peter Höglund1
1Department of Clinical Sciences, Lund University, Lund, Sweden.
Background:
Computed tomographic angiography (CTA) may require a non-contrast enhanced dataset for the diagnostic workup. By using dual-energy acquisition, it is possible to obtain a virtual non-contrast-enhanced (VNC) dataset, thereby possibly eliminating the non-contrast acquisition and reducing the radiation dose.
Purpose:
To compare image quality of VNC images reconstructed from arterial phase dual-energy CTA to true non-contrast (TNC) images, and to assess whether VNC images were of sufficient quality to replace TNC images.
Material And Methods:
Thirty consecutive patients with suspected abdominal aortic aneurysm, aortic dissection, or subacute control after EVAR/TEVAR were examined with dual-energy CT (DECT). The examination protocol included a single-energy TNC, DECT arterial phase (80 kV/Sn140 kV), and single-energy in venous phase of the aorta. A VNC dataset was obtained from the DE acquisition from arterial phase scans. Mean attenuation and image noise were measured within regions of interest at three levels in the aorta in TNC and VNC images. Comparison of the TNC and VNC images for artefacts was made side-by-side. Subjective evaluation included overall image quality on a 4-grade scale, and quantitative analysis of algorithm-induced artefacts by two experienced readers.
Results:
For all cases, the aortic attenuation was significantly higher at VNC than at TNC. Image noise measured quantitatively was also significantly higher at VNC than at TNC. Subjective image quality was lower for VNC (mean = 3.1 for VNC, 3.7 = for TNC) but there were no cases rated non-diagnostic.
Conclusion:
VNC images based on arterial phase CTA have significantly higher mean attenuation and higher noise levels compared to TNC.
More Related Videos
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
04:40Identifying Coronary Artery Calcification on Non-gated Computed Tomography Scans
Published on: August 28, 2018
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Reliability and Validity
Virtual Work
In static equilibrium, a body can experience an imaginary or virtual movement, such as displacement or rotation. The virtual work done by a force is equal to the dot product of force and virtual displacement in the direction of the force. When it comes to virtually rotating a...
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Distribution Reliability and Automation