Ambient dose during intra-oral radiography with current techniques: Part 2 quantifying the remnant beam - an in vivo
Reinier C Hoogeveen1, Christine Ch Pfältzer1, Wer Berkhout1
11 Department of Oral and Maxillofacial Radiology, Academic Centre for Dentistry Amsterdam (ACTA), University of Amsterdam and VU University Amsterdam , Amsterdam , The Netherlands.
Objectives::
(1) To find a safe reference transmission factor for the remnant beam to be used when assessing the remnant beam during intra-oral radiography. (2) To identify factors such as exposure type, side (left or right), gender and age that might significantly affect transmission.
Methods::
Measurements were performed in the remnant beam during 323 intra-oral exposures of 43 patients. The exposure parameters were 60 kV and 7 mA. These measurements were compared with values in the same setting, but without a patient present to arrive at a transmission value. Differences between types of exposure [bitewing (BW), incisor and canine (IC) periapical, premolar and molar (PM) periapical], exposure side (left or right), gender and age were statistically analysed. The reference value was based on the exposure yielding the highest transmission value, to which a safety margin of two standard deviations (SDs) was added, and then rounding up.
Results::
The respective mean transmission values (SD) for BW, IC and PM exposures were as follows: 1.78% (1.15%), 0.639% (0.63%) and 2.60% (0.98%). The differences between PM and IC and between BW and IC were significant (p > 0.01). The differences between exposure side, gender and age were not significant. The reference transmission value, which was calculated from the highest mean transmission percentage for PM (2.60%) plus twice the SD (0.98%) and rounding up, was 5%.
Conclusions::
A transmission factor of 5% could be used to assess the remnant beam during intra-oral radiography. Anterior exposures led to significantly lower transmission values than posterior exposures.
Related Concept Videos
Bioavailability Study Design: Single Versus Multiple Dose Studies
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
Deflection of a Beam
Singularity functions, described in an earlier lesson, are powerful mathematical tools that represent discontinuities within a function commonly encountered in structural loading...
Oral Cavity
Teeth: The teeth are the hardest structures in our bodies. Humans have two sets of teeth throughout their lifetime: deciduous (baby) teeth and permanent teeth. Each tooth consists of several parts: the crown (visible part), the root (embedded in the jaw), enamel (hard outer...
Dose Size and Dosing Frequency: Determination Methods
Quantifying Work