Optical coherence tomography-verified longer balloon inflation time may provide better stent apposition and optimal
O Tasar1,2, A K Karabay3, C Yücel Karabay4
1Elazig Research and Education Hospital, Elazig, Turkey. tasaronur@hotmail.com.
Background:
Incomplete stent expansion and inadequate apposition predispose to stent thrombosis following percutaneous coronary intervention. Recent studies have shown that increasing the duration of balloon inflation during stent employment was beneficial. Thus, the balloon inflation time required for optimal stent expansion and apposition in patients receiving second-generation drug-eluting stents (DES) were determined using optical coherence tomography (OCT).
Patients And Methods:
Between April 2014 and March 2015, 38 patients (28 men, 10 women; mean age 60.5 ± 11.4 years) with stable angina pectoris due to single significant de novo coronary artery stenosis were prospectively enrolled. All patients were administered aspirin and clopidogrel and received weight-adjusted intravenous unfractionated heparin. Images of basal lesions were obtained using the C7XR LightLab Dragonfly OCT catheter.
Results:
Expansion and apposition parameters improved with increasing duration of balloon inflation (30 s or 60 s) with nominal pressure (12 atm). Mean lesion length was 19.8 ± 7.6 mm. Mean stent diameter and length were 2.8 ± 0.36 mm and 24.9 ± 7.6 mm, respectively.
Conclusion:
With deployment of a stent at nominal pressure with conventional duration, inadequate stent expansion and malapposition frequently occurred as detected by OCT; however, a balloon inflation duration of 60 s markedly improved stent expansion and apposition parameters without significant complications.
Related Concept Videos
Wave Parameters
Optimal Foraging
Optimization Problems
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Distributions to Estimate Population Parameter
Noncompartmental Analysis: Miscellaneous Pharmacokinetic Parameters
One key aspect of the noncompartmental approach is determining a drug's total clearance. This can be done by dividing the drug dose by the area under the concentration-time curve from zero to infinity. The area under the concentration-time curve represents the drug's...