Immunoluminometric assay for copeptin measurement in cerebrospinal fluid: Technical aspects and pilot study
S Mindt1, H Andrade-Barazarte2, U Tokhi2
1Institut for Clinical Chemistry, University Hospital Mannheim, Theodor-Kutzer-Ufer 1-3, Medical Faculty University of Heidelberg, 68167 Mannheim, Germany.
Background:
Copeptin acts as surrogate marker under stress stimuli, as well as an outcome predictor based on serum or plasma concentration in patients suffering intracranial hemorrhage, aneurysmal subarachnoid hemorrhage (aSAH), and stroke. The aim of this study was to establish a method for quantification of copeptin levels in cerebrospinal fluid (CSF) and to demonstrate its clinical applicability in patients following aSAH.
Methods:
This assay was validated for CSF samples using a commercial immunoluminometric assay (IMLA). For the control group (10 patients), CSF copeptin levels were determined in patients without signs of acute neurological diseases and who underwent a diagnostic lumbar puncture. The pilot cohort included calculation of copeptin levels in CSF and in serum of patients following aSAH.
Results:
The control group had CSF copeptin levels lower than 0.78 pmol/L-1. Among patients with aSAH, CSF copeptin values had a mean of 20.1 pmol/L-1 and serum copeptin concentrations had a mean of 61.39 pmol/L-1.
Conclusions:
This assay provides to best of our knowledge for the first time initial ranges values of CSF copeptin for patients without acute neurological disease and in patients with aSAH. Thus, it opens new doors to develop further calculations and relationships between diseases biomarker and outcome prediction.
Related Concept Videos
Cerebrospinal Fluid
CSF Production
CSF is produced mainly in the choroid plexus, a network of capillaries and ependymal cells located within the ventricular system of the brain....
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
Pilot and Numeric Relaying
Quantitative Aspects of Drug-Receptor Interaction
The Fluid Mosaic Model
Fluid Pressure
According to Pascal's law, a fluid at rest will generate equal pressure in all directions. This pressure is measured as a force per unit area, and its magnitude depends on the fluid's specific...