The Varian Halcyon linear accelerator is designed for faster treatment time and fewer treatment errors than conventional linear accelerators. Its increased gantry speed (4 rpm), dose rate (800 MU/min), and leaf speed (5 cm/s) reduce treatment time. The 2 multileaf collimator banks of 1 cm leaf width could cause planning difficulties, but utilizing many arcs could address this challenge. While 3 or more arcs are often avoided due to the increased treatment time, the Halcyon can deliver 5 arcs in under 3 minutes. For prostate and head-and-neck, we hypothesized plan quality would increase with increasing arcs without compromising treatment time. Fifteen prostate and fifteen head-and-neck previously treated patients were replanned to 78 and 66 Gy, respectively. A total of 150 volumetric modulated arc therapy (VMAT) plans, ranging from 1 to 5 arcs, were created to compare organ-at-risk (OAR) sparing, target coverage, homogeneity, and hotspot volume. Plans were normalized to the prescribed dose of 98% and 95% of the primary target volume for prostate and head-and-neck, respectively. All prostate 4-arc and 5-arc plans met MD Anderson Cancer Center (MDACC) constraints, while 86.7%, 20%, and 6.7% of the 1-arc, 2-arc, and 3-arc plans did not meet constraints, respectively. As the number of arcs increased, all OAR maximum doses decreased. All head-and-neck 5-arc plans met MDACC constraints, while 73.3%, 26.7%, 13.3%, and 6.7% of the 1-arc, 2-arc, 3-arc, and 4-arc plans did not meet constraints, respectively. The contralateral parotid mean dose and the spinal cord maximum dose decreased with increasing arcs. Homogeneity and conformity indices decreased when increasing from 1 to 5 arcs for both prostate and head-and-neck. Overall, homogeneity and conformity indices and OAR doses improved when using three or more arcs. Therefore, using at least 3 arcs for VMAT plans on the Halcyon will increase plan quality without compromising treatment time.
Related Concept Videos
Arteries of the Head and Neck
The internal carotid arteries supply blood to the anterior portion of the cerebrum. They enter the...
Veins of Head and Neck
On the other hand, the vertebral veins, unlike their arterial counterparts, are not primarily responsible for brain drainage. Instead, they drain the cervical vertebrae, spinal cord, and some small...
Muscles of the Anterior Neck
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Muscles that Move the Head
The bilateral sternocleidomastoid, or SCM, and the suprahyoid and infrahyoid muscles are significant head flexors. The SCM muscles originate at the sternum and clavicle and attach to the mastoid process of the temporal bone. The SCM contracts bilaterally to bend the head forward, whereas...
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...