Accuracy of frame-based and frameless systems for deep brain stimulation: A meta-analysis
Andrew Roth1, Simon S Buttrick1, Iahn Cajigas1
1Department of Neurological Surgery University of Miami/Jackson Memorial Hospital, Miami, FL, United States.
Summary
Frameless systems for deep brain stimulation (DBS) offer accuracy comparable to traditional frame-based stereotaxy. While frame-based methods show slight statistical benefits, the small differences suggest frameless approaches are a viable alternative for movement disorder treatments.
Area of Science:
- Neurosurgery
- Neurology
- Medical Engineering
Background:
- Deep brain stimulation (DBS) is a key treatment for movement disorders, requiring precise lead placement in the basal ganglia.
- Frame-based stereotaxy has been the traditional standard for accurate DBS lead implantation.
- Emerging frameless systems aim to provide comparable accuracy without the use of a stereotactic frame.
Purpose of the Study:
- To conduct a meta-analysis comparing the targeting accuracy of frameless versus frame-based stereotactic systems for DBS lead placement.
- To quantitatively assess the differences in electrode placement error across cardinal directions (x, y, z) between the two systems.
Main Methods:
- Systematic literature search identifying studies reporting DBS targeting accuracy for both frameless and frame-based methods.
- Inclusion of 5 relevant studies after screening 76 papers, evaluating a total of 425 DBS electrode placements.
- Statistical analysis using standard difference of means to compare targeting errors between frameless and frame-based stereotaxy.
Main Results:
- Frame-based stereotaxy demonstrated statistically significant accuracy benefits in the x (p=0.036) and y (p=0.0025) coordinates.
- No significant difference in targeting accuracy was observed for the z coordinate between the two methods.
- Composite mean differences in error were minimal: 0.3037 mm (x), 0.0305 mm (y), and 0.1630 mm (z).
Conclusions:
- Frameless stereotactic systems are a reasonable alternative to frame-based methods for DBS surgery.
- Although statistically significant accuracy differences exist favoring frame-based systems, the magnitude of these differences is very small.
- The observed small discrepancies in accuracy are likely of questionable clinical significance, supporting the use of frameless DBS techniques.