Nitrogen Compound Characterization in Fuels by Multidimensional Gas Chromatography
Published on: May 15, 2020
Ultrafast separations via pulse flow valve modulation to enable high peak capacity multidimensional gas
H Daniel Bahaghighat1, Chris E Freye2, Derrick V Gough2
1Department of Chemistry, Box 351700, University of Washington, Seattle, WA 98195, USA; Department of Chemistry and Life Science, United State Military Academy, West Point, NY 10996, USA.
Abstract:
Ultrafast modulation with a modulation period PM ≥ 50ms via a pulse flow valve is demonstrated for comprehensive two-dimensional gas chromatography (GC×GC) and comprehensive three-dimensional (3D) gas chromatography (GC3). Significant increases in peak capacity and peak capacity production are achieved for GC×GC and GC3 relative to previous studies due to using pulse flow valve modulation. Due to the nature of the "partial" modulation process, the separation dimension following pulse flow valve modulation is not a traditional chromatogram, rather requires data processing to convert the data to expose the encoded chromatographic information, producing "apparent" chromatographic peaks. In the GC×GC mode, a 115-component test mixture was evaluated using a PM of 500ms, creating an apparent 2D peak width-at-base 2W with an average of 25ms, producing a 2nc of 20. Based on the average 1W of 1.0s for the 6min first dimension 1D separation, an ideal peak capacity nc,2D of 7200 is achieved (1,200/min peak production). For a high-speed GC×GC separation (30s run), a PM of 75ms produced apparent 2W of 8ms, ideal for the third dimension of a GC3 instrument. Using the knowledge gained from this high-speed GC×GC experiment, the pulse flow valve was implemented as the second modulator in GC3. Three samples were evaluated in the GC3 mode: a simple mixture containing 18 compounds (to illustrate basic concepts), the 115-component test mixture (to determine peak capacity figures-of-merit), and a diesel spiked with 8 polar compounds (to illustrate chemical selectivity benefits of GC3). For the 115-component test mixture with a 1PM of 1.2s and a 2PM of 60ms, average 1W of 3.2s, 2W of 130ms, and apparent 3W of 13ms were produced, resulting in a 1nc of 210, 2nc of 9.2, and 3nc of 5, respectively. Hence, an ideal peak capacity, nc,3D of ∼10,000 for GC3 was achieved for the 11min 1D separation window of the 115-component test mixture.
Related Concept Videos
Heat Capacities of an Ideal Gas I
Molar heat capacity quantifies the ratio of the amount of heat added (or removed) to increase (or decrease) the...
Heat Capacities of an Ideal Gas II
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...
Heat Capacities of an Ideal Gas III
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...