Deep Vascular Imaging in the Eye with Flow-Enhanced Ultrasound
Published on: October 4, 2021
Ultrasound Imaging and Measurement of Choroidal Blood Flow
Raksha Urs1, Jeffrey A Ketterling2, Alfred C H Yu3
1Department of Ophthalmology, Columbia University Medical Center, New York, NY, USA.
Purpose:
The choroid is a vascular network providing the bulk of the oxygen and nutrient supply to the retina and may play a pivotal role in retinal disease pathogenesis. While optical coherence tomography angiography provides an en face depiction of the choroidal vasculature, it does not reveal flow dynamics. In this report, we describe the use of plane-wave ultrasound to image and characterize choroidal blood flow.
Methods:
We scanned both eyes of 12 healthy subjects in a horizontal plane superior to the optic nerve head using an 18-MHz linear array. Plane-wave data were acquired over 10 transmission angles that were coherently compounded to produce 1000 images/sec for 3 seconds. These data were processed to produce a time series of power Doppler images and spectrograms depicting choroidal flow velocity. Analysis of variance was used to characterize peak systolic, and end diastolic velocities and resistive index, and their variability between scans, eyes, and subjects.
Results:
Power Doppler images showed distinct arterioles within a more diffuse background. Choroidal flow was moderately pulsatile, with peak systolic velocity averaging approximately 10 mm/sec and resistive index of 0.55. There was no significant difference between left and right eyes, but significant variation among subjects.
Conclusions:
Plane-wave ultrasound visualized individual arterioles and allowed measurement of flow over the cardiac cycle. Characterization of choroidal flow dynamics offers a novel means for assessment of the choroid's role in ocular disease.
Translational Relevance:
Characterization of choroidal flow dynamics offers a novel means for assessment of the choroid's role in ocular disease.
Related Concept Videos
Blood Flow
Autoregulation of Blood Flow
Chemical Signaling in Autoregulation
Chemical signaling operates at the precapillary sphincter level, inciting either contraction or relaxation....
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Applications of Integration to Find Blood Flow
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...