Determination of the dose rate constant through Monte Carlo simulations with voxel phantoms
Abner D Soares1, Lucas Paixão2, Alessandro Facure3
1Instituto de Radioproteção e Dosimetria, Avenida Salvador Allende, 9, 22780-160, Rio de Janeiro, RJ, Brazil.
Purpose:
The main goal of this study was to determine the dose rate constant of radionuclides used in Nuclear Medicine when they are biodistributed in humans. The secondary goal was to calculate the effective half-life for the same radionuclides in order to determine the realistic dose due to the incorporation of these nuclides for a variety of reasons.
Methods:
Anthropomorphic voxel phantoms, with characteristics based on ICRP-110 were considered and the biodistributions of radionuclides were simulated using the Monte Carlo code MCNPX. In addition, the effective half-lives for all simulated radionuclides were calculated.
Results:
The results for the dose rate constant, taking into account the patient body attenuation, showed no correlation between the biodistribution of the radionuclide and the energy of the radiation emitted. Body attenuation ranged between 6.7% and 94.1%. Also, the priority angle of radiation emission was determined. The values found for the body attenuation agree with the literature, corroborating the calculation of the body dose rate constant.
Conclusions:
The results for the body dose rate constant calculated in this work, for 30 radionuclides and 57 radiopharmaceuticals, and their effective half-lives, may be used to estimate the dose emitted by a person who has incorporated a radionuclide in the Nuclear Medicine activity range. This dose will be optimized when the body dose rate constant and the effective half-life determined in this study are used together with dose reduction factors.
More Related Videos
Related Concept Videos
Determination of Michaelis Constant and Maximum Elimination Rate
These parameters can be estimated by analyzing plasma concentration data post-drug administration. A notable example of this application is phenytoin, a drug with capacity-limited kinetics. It's recommended that phenytoin should be administered at two...
Rate-Determining Steps
In a multistep reaction mechanism, one of the elementary steps progresses significantly slower than the others. This slowest step is called the rate-limiting step (or rate-determining step). A reaction cannot proceed faster than its slowest step, and hence, the rate-determining step limits the overall reaction rate.
The concept of rate-determining step can be understood from the analogy of a 4-lane freeway with a short-stretch of traffic-bottleneck caused due to...
Drug Dosing in Renal Diseases: Dose Adjustments Based on Drug Clearance and Elimination Rate Constant
Dose Size and Dosing Frequency: Determination Methods
Determination of Multiple Dosing Parameters: Loading and Maintenance Doses
Calculating the Equilibrium Constant
For example, gaseous nitrogen dioxide forms dinitrogen tetroxide according to this equation: