Planning comparison of five automated treatment planning solutions for locally advanced head and neck cancer
J Krayenbuehl1, M Zamburlini2, S Ghandour3
1Department of Radiation Oncology, University Hospital Zurich, Rämistrasse 100, CH-8091, Zurich, Switzerland. Jerome.krayenbuehl@usz.ch.
Background:
Automated treatment planning and/or optimization systems (ATPS) are in the process of broad clinical implementation aiming at reducing inter-planner variability, reducing the planning time allocated for the optimization process and improving plan quality. Five different ATPS used clinically were evaluated for advanced head and neck cancer (HNC).
Methods:
Three radiation oncology departments compared 5 different ATPS: 1) Automatic Interactive Optimizer (AIO) in combination with RapidArc (in-house developed and Varian Medical Systems); 2) Auto-Planning (AP) (Philips Radiation Oncology Systems); 3) RapidPlan version 13.6 (RP1) with HNC model from University Hospital A (Varian Medical Systems, Palo Alto, USA); 4) RapidPlan version 13.7 (RP2) combined with scripting for automated setup of fields with HNC model from University Hospital B; 5) Raystation multicriteria optimization algorithm version 5 (RS) (Laboratories AB, Stockholm, Sweden). Eight randomly selected HNC cases from institution A and 8 from institution B were used. PTV coverage, mean and maximum dose to the organs at risk and effective planning time were compared. Ranking was done based on 3 Gy increments for the parallel organs.
Results:
All planning systems achieved the hard dose constraints for the PTVs and serial organs for all patients. Overall, AP achieved the best ranking for the parallel organs followed by RS, AIO, RP2 and RP1. The oral cavity mean dose was the lowest for RS (31.3 ± 17.6 Gy), followed by AP (33.8 ± 17.8 Gy), RP1 (34.1 ± 16.7 Gy), AIO (36.1 ± 16.8 Gy) and RP2 (36.3 ± 16.2 Gy). The submandibular glands mean dose was 33.6 ± 10.8 Gy (AP), 35.2 ± 8.4 Gy (AIO), 35.5 ± 9.3 Gy (RP2), 36.9 ± 7.6 Gy (RS) and 38.2 ± 7.0 Gy (RP1). The average effective planning working time was substantially different between the five ATPS (in minutes): < 2 ± 1 for AIO and RP2, 5 ± 1 for AP, 15 ± 2 for RP1 and 340 ± 48 for RS, respectively.
Conclusions:
All ATPS were able to achieve all planning DVH constraints and the effective working time was kept bellow 20 min for each ATPS except for RS. For the parallel organs, AP performed the best, although the differences were small.
Related Concept Videos
Sampling Plans
Random sampling is a method where each member of the population has an equal chance of being selected for the sample. It involves selecting individuals randomly, often using random number generators or lottery-type methods. For example, when analyzing the properties of a...
Planning Nursing Care I
Planning Nursing Care II
Arteries of the Head and Neck
The internal carotid arteries supply blood to the anterior portion of the cerebrum. They enter the...
Veins of Head and Neck
On the other hand, the vertebral veins, unlike their arterial counterparts, are not primarily responsible for brain drainage. Instead, they drain the cervical vertebrae, spinal cord, and some small...
Role of Communication in the Nursing Process II: Planning and Implementation