Intraoperative performance and ultrastructural integrity of human capsulotomies created by the improved precision
Joobin Hooshmand1, Robin G Abell1, Penny Allen1
1From the Tasmanian Eye Institute (Hooshmand, Abell, Allen, Vote), Launceston, Tasmania, Royal Victorian Eye and Ear Hospital (Abell), Melbourne, Victoria, Rural Clinical School (Allen), University of Tasmania, Launceston, Tasmania, and Central Science Laboratory (Goemann), University of Tasmania, Hobart, Tasmania, Australia.
Purpose:
To investigate the intraoperative performance and the ultrastructural features of anterior capsulotomy performed with the improved Zepto precision pulse capsulotomy (PPC) device.
Setting:
Launceston Eye Institute, Launceston, Tasmania, Australia.
Design:
Prospective, consecutive case series.
Methods:
Intraoperative observation of performance and complications were made on consecutive eyes having capsulotomies with the improved PPC device. The capsulotomy specimens were examined under scanning electron microscope and compared with manual continuous curvilinear capsulorhexis and femtosecond laser-assisted cataract surgery specimens.
Results:
The study comprised 52 eyes. Intended capsulotomy occurred in 50 eyes with 2 device failures. Complete free-floating capsulotomies were achieved in 48 eyes (96%). Anterior capsule tears occurred in 2 eyes (4%).
Conclusions:
The ability of the PPC device to create complete free-floating capsulotomies improved; however, the radial tear rate remained high. The ultrastructural features in ex vivo human capsulotomy specimens still showed areas of irregular capsule margin with frayed edges.
More Related Videos
06:16A Probing Device for Quantitatively Measuring the Mechanical Properties of Soft Tissues during Arthroscopy
Published on: May 1, 2020
05:56Evaluation of the Feasibility, Safety, and Accuracy of an Intraoperative High-intensity Focused Ultrasound Device for Treating Liver Metastases
Published on: January 9, 2019
Related Concept Videos
Pulse
The pulse serves as a clinical...
Pulse
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls...
Uncertainty in Measurement: Accuracy and Precision
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Regulation of Pulse