A Monte-Carlo-based and GPU-accelerated 4D-dose calculator for a pencil beam scanning proton therapy system
Mark D Pepin1, Erik Tryggestad1, Hok Seum Wan Chan Tseung1
1Department of Radiation Oncology, Mayo Clinic, 200 1st Street Southwest, Rochester, MN, 55905, USA.
Purpose:
The presence of respiratory motion during radiation treatment leads to degradation of the expected dose distribution, both for target coverage and healthy tissue sparing, particularly for techniques like pencil beam scanning proton therapy which have dynamic delivery systems. While tools exist to estimate this degraded four-dimensional (4D) dose, they typically have one or more deficiencies such as not including the particular effects from a dynamic delivery, using analytical dose calculations, and/or using nonphysical dose-accumulation methods. This work presents a clinically useful 4D-dose calculator that addresses each of these shortcomings.
Methods:
To quickly compute the 4D dose, the three main tasks of the calculator were run on graphics processing units (GPUs). These tasks were (a) simulating the delivery of the plan using measured delivery parameters to distribute the plan amongst 4DCT phases characterizing the patient breathing, (b) using an in-house Monte Carlo simulation (MC) dose calculator to determine the dose delivered to each breathing phase, and (c) accumulating the doses from the various breathing phases onto a single phase for evaluation. The accumulation was performed by individually transferring the energy and mass of dose-grid subvoxels, a technique that models the transfer of dose in a more physically realistic manner. The calculator was run on three test cases, with lung, esophagus, and liver targets, respectively, to assess the various uncertainties in the beam delivery simulation as well as to characterize the dose-accumulation technique.
Results:
Four-dimensional doses were successfully computed for the three test cases with computation times ranging from 4-6 min on a server with eight NVIDIA Titan X graphics cards; the most time-consuming component was the MC dose engine. The subvoxel-based dose-accumulation technique produced stable 4D-dose distributions at subvoxel scales of 0.5-1.0 mm without impairing the total computation time. The uncertainties in the beam delivery simulation led to moderate variations of the dose-volume histograms for these cases; the variations were reduced by implementing repainting or phase-gating motion mitigation techniques in the calculator.
Conclusions:
A MC-based and GPU-accelerated 4D-dose calculator was developed to estimate the effects of respiratory motion on pencil beam scanning proton therapy treatments. After future validation, the calculator could be used to assess treatment plans and its quick runtime would make it easily usable in a future 4D-robust optimization system.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
07:07Author Spotlight: High-Resolution 4D Light-Sheet Imaging and Virtual Reality in Zebrafish for Single-Cell Analysis of Heart Function
Published on: January 5, 2024
Related Concept Videos
Titration Calculations: Weak Acid - Strong Base
For the titration of 25.00 mL of 0.100 M CH3CO2H with 0.100 M NaOH, the reaction can be represented as:
Titration Calculations: Strong Acid - Strong Base
A titration is carried out for 25.00 mL of 0.100 M HCl (strong acid) with 0.100 M of a strong base NaOH. The pH at different volumes of added base solution can be calculated as follows:
(a) Titrant volume = 0 mL. The solution pH is due to the acid ionization of HCl. Because this is a strong acid, the ionization is complete and the hydronium ion molarity is 0.100 M. The pH of the solution is then:
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
Calculating pH Changes in a Buffer Solution
Accelerators
The effectiveness of calcium chloride can...
Drug Dosing in Renal Diseases: Dose Adjustments Based on Drug Clearance and Elimination Rate Constant