A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Avoiding inappropriate therapy of single-lead implantable cardioverter-defibrillator by using atrial-sensing
Muhammed Kurt1, Nikesh Jathanna1, Mehran Babady1
1Department of Cardiology, Pulmonology, and Vascular Medicine, Medical Faculty of the Heinrich Heine University Düsseldorf, Düsseldorf, Germany.
Introduction:
The single-chamber implantable cardioverter-defibrillator (ICD) can be associated with more frequent inappropriate therapies compared with dual-chamber ICDs, when they are accompanied by a simpler implantation procedure. The aim of this study was to investigate whether the use of a single-lead ICD system with atrial-sensing electrodes results in a reduction of inappropriate ICD therapy.
Methods And Results:
The study population consisted of 212 consecutive patients, who underwent primary prophylactic single-lead ICD implantation at our institute. A ventricular lead with atrial-sensing electrodes was implanted in 77 patients (36%; Group-VDD) and a ventricular lead without atrial-sensing electrodes was implanted in 135 patients (64%; Group-VVI). Procedural and follow-up data were collected in a prospective registry. A higher prevalence of atrial fibrillation was present in Group-VDD. There were no other significant differences in patient baseline characteristics (age, sex, and other comorbidities) or follow-up period between the two groups. The operative parameters including fluoroscopic burden showed no significant differences between Group-VDD and Group-VVI. During a mean follow-up period of 697 ± 392 days, 26 patients (12%) experienced appropriate ICD therapies and 13 patients (6%) suffered inappropriate ICD therapies. The incidence of inappropriate ICD therapies in Group-VDD was significantly lower as compared to that of Group-VVI (1/77 [1%] vs 12/135 [9%]; log-rank, P = 0.028). The incidence of appropriate ICD therapies and the occurrence of device-related complications showed no significant difference between the two groups.
Conclusion:
Single-lead ICD with atrial-sensing electrodes shows a lower incidence of inappropriate ICD therapy compared with the absence of atrial-sensing electrodes, without additional operative burden or increased complications.
Related Concept Videos
Frustration and Conflict: Avoidance-Avoidance, Double-Approach Avoidance
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Standard Electrode Potentials
Introduction to Special Senses
Personality Disorders: Narcissistic and Avoidant
Characteristics of Narcissistic Personality Disorder
Narcissistic individuals exhibit an inflated sense of self-importance and an excessive need for admiration. They are often...
Avoidance Learning and Learned Helplessness
Avoidance learning occurs when an organism learns that a specific behavior can prevent an unpleasant outcome. For example, a student who receives a bad grade may start studying harder to avoid future poor grades. This behavior persists even when the negative outcome is no longer present. Avoidance learning is powerful because it maintains behavior in the absence of the...