Impact of prompt gamma coincidence correction on absorbed dose estimation in differentiated thyroid cancer using 124I
Roel Wierts1, Maurizio Conti2, Anouk G G Claessen1
1Department of Radiology and Nuclear Medicine, Maastricht University Medical Center, Maastricht, The Netherlands.
Introduction:
Iodine-124 positron emission tomography/computed tomography (I PET/CT) is increasingly being used in the absorbed dose estimation in the radioiodine treatment of differentiated thyroid cancer (DTC). However, the produced prompt gamma coincidences (PGCs) associated with the I decay result in a bias in the absorbed dose estimation. The impact of a sinogram-based PGC correction approach on the absorbed dose estimation in I PET/CT DTC imaging is investigated.
Methods:
I phantom and patient measurements were performed on a Siemens Biograph mCT PET/CT system. All images were reconstructed with (PGCon) and without PGC correction (PGCoff). The phantom contained seven spheres (diameters: 6.6-37 mm). The spheres and background compartment were filled with a I solution, resulting in a low (9.4 : 1) and a high sphere-to-background activity concentration ratio (750 : 1). Sphere recovery coefficient (RC) values were determined. In addition, the impact of PGC correction on measured lesion uptake and calculated lesion-absorbed dose was assessed for 66 lesions identified in 24 DTC patients.
Results:
PGC correction systematically increased sphere RC values up to 71% for the smallest spheres. For the patient data, PGC correction significantly increased both the measured I uptake (P<0.005) and the calculated lesion-absorbed dose (P=0.008) by ∼3%. The percentage difference in the calculated lesion-absorbed dose ranged from -19% to 50%, showing that PGC correction had a variable and large impact for a few lesions.
Conclusion:
PGC correction resulted in significantly higher sphere RC values, I lesion uptake values and estimated lesion-absorbed doses.
More Related Videos
07:54Implantation and Monitoring by PET/CT of an Orthotopic Model of Human Pleural Mesothelioma in Athymic Mice
Published on: December 21, 2019
06:51Utilizing 18F-FDG PET/CT Imaging and Quantitative Histology to Measure Dynamic Changes in the Glucose Metabolism in Mouse Models of Lung Cancer
Published on: July 21, 2018
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Distance Corrections
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Power Factor Correction
The Thyroid Gland
The follicles have a central cavity lined by simple cuboidal to squamous epithelial cells called follicular cells. These cells produce the glycoprotein...