Myocardial Thermal Ablation with a Transesophageal High-Intensity Focused Ultrasound Probe: Experiments on Beating
Paul Greillier1, Bénédicte Ankou2, Pierre Bour3
1Universite Lyon, INSERM, UMR1032, LabTAU, Lyon, France.
Abstract:
Described here is a study of transesophageal thermal ablation of isolated and perfused beating hearts and non-human primates. An endoscope integrating a transesophageal echocardiography probe and a high-intensity focused ultrasound transducer was built and tested on five Langendorff-isolated hearts and three 30-kg baboons. B-Mode ultrasound, passive elastography and magnetic resonance imaging were performed to monitor thermal lesions. In isolated hearts, continuous and gated sonication parameters were evaluated with acoustic intensities of 9-12 W/cm2. Sonication parameters of gated exposures with 12 W/cm2 acoustic intensity for 5 min consistently produced visible lesions in the ventricles of isolated hearts. In animals, left atria and ventricles were exposed to repeated continuous sonications (4-15 times for 16 s) at an acoustic intensity at the surface of the transducer of 9 W/cm2. Clinical states of the baboons during and after the treatment were good. One suspected lesion in the left ventricle could be evidenced by elastography, but was not confirmed by magnetic resonance imaging. The transesophageal procedure therefore has the potential to create thermal lesions in beating hearts and its safety in clinical practice seems promising. However, further technical exploration of the energy deposition in the target would be necessary before the next pre-clinical experiments.
More Related Videos
05:56Evaluation of the Feasibility, Safety, and Accuracy of an Intraoperative High-intensity Focused Ultrasound Device for Treating Liver Metastases
Published on: January 9, 2019
09:03Disruption of the Blood-Spinal Cord Barrier Using Low-Intensity Focused Ultrasound in a Rat Model
Published on: March 10, 2023
Related Concept Videos
Beats
Sound Intensity
Thermal expansion and Thermal stress: Problem Solving
To solve the problem, first, identify the known and unknown quantities. The initial length (L) of the bridge is 1275 m, the coefficient of linear expansion (α) for steel is 12 x 10-6/°C, and the change in temperature (ΔT) is 55...
Sound Intensity Level
The human ear can perceive an extensive range of sound intensity, necessitating the use of the logarithmic scale to define a physical quantity—the intensity level. It is a ratio of two intensities and...
What is an Experiment?
Thermal Strain