Comparison of measured and calculated doses in a Rando phantom with a realistic lung radiotherapy treatment plan
Serine Sarra Bouacid1,2, Fayçal Kharfi3,4, Fouad Boulakhssaim5
1Department of Physics, University of Ferhat Abbas-Setif1, Campus El-Bèz, 19000, Sétif, Algeria.
Abstract:
In this work, dose measurements were performed to evaluate an external radiotherapy treatment plan and, particularly, to validate dose calculations for a lung lesion case. Doses were calculated by the Varian Eclipse treatment planning system using the AAA anisotropic analytical algorithm. The measurements were performed using a Rando anthropomorphic phantom and TLD700 thermoluminescent dosimeters. The comparison between doses calculated and doses measured by means of thermoluminescence (TL) shows compatibility except for a few points, due to the limitations in the heterogeneity correction used for the case studied here. The deviation between the calculated and measured doses is about 6.5% for low (< 0.5 Gy) doses and about 1% for higher doses (> 0.5 Gy).The deviation between AAA-calculated and TL-measured doses was also found to be higher in proximity to heterogeneous tissue interfaces.
Related Concept Videos
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Lung Capacity
Calculating the Equilibrium Constant
For example, gaseous nitrogen dioxide forms dinitrogen tetroxide according to this equation:
Sampling Plans
Random sampling is a method where each member of the population has an equal chance of being selected for the sample. It involves selecting individuals randomly, often using random number generators or lottery-type methods. For example, when analyzing the properties of a...
Calculating Standard Free Energy Changes
Calculating pH Changes in a Buffer Solution