Variability in radiation dose and image quality: A comparison across fluoroscopy-system vendors, generations of
Kevin D Hill1, Steve D Mann2, Michael P Carboni1
1Division of Pediatric Cardiology, Department of Pediatrics, Duke University Medical Center, Durham, North Carolina.
Objectives:
To evaluate differences in radiation dose and image quality across institutions, fluoroscope vendors and generations of fluoroscopes for pediatric cardiac catheterization.
Background:
Increased recognition of the potentially harmful effects of ionizing radiation has spurred technological advances in fluoroscopes, as well as increased focus on optimizing fluoroscope performance. There is currently little understanding of variability in the dose-image quality relationship across institutions, fluoroscope vendor and/or generation of equipment.
Methods:
We evaluated latest generation fluoroscopes from Phillips, Siemens, GE, and Toshiba, and an older generation Phillips fluoroscope (release date 2003) at three different institutions. Radiation dose was measured using an anthropomorphic dose-assessment phantom with effective dose in mSv estimated from Monte Carlo simulations. Image quality phantom images were scored on a 12-point scale by three blinded reviewers.
Results:
Fluoroscope effective doses ranged from 0.04 to 0.14 mSv/1,000 pulses for fluoroscopy with associated composite image quality scores ranging from 8.0 ± 0.6 to 10.4 ± 1.3. For cineangiography, effective doses ranged from 0.17 to 0.57 mSv/1,000 frames with image quality scores ranging from 10.1 ± 0.3 to 11.1 ± 0.3. There was modest correlation between effective dose and image quality (r = 0.67, P = 0.006). The older generation fluoroscope delivered consistently higher doses than the newer generation systems (2.3- to 3.5-fold higher for fluoroscopy; 1.1- to 3.4-fold higher for cineangiography) without appreciable differences in image quality.
Conclusion:
Technological advances have markedly improved fluoroscope performance. Comparing latest generation systems across vendors and institutions, we found variability in the dose-IQ relationship and speculate that this reflects both equipment and institutional optimization practices.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Generating Electromagnetic Radiations
Biological Effects of Radiation
Leveling Equipment
Personal Protective Equipment
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Radiation: Applications
The average...