Related Experiment Video
Updated: Feb 5, 2026

Preparation of Rat Tail Tendons for Biomechanical and Mechanobiological Studies
Published on: July 30, 2010
Attenuation of Proximal Junctional Kyphosis Using Sublaminar Polyester Tension Bands: A Biomechanical Study
Samuel K Cho1, John Caridi1, Jun S Kim1
1Department of Orthopedic Surgery, Icahn School of Medicine at Mount Sinai, New York, New York, USA.
Objective:
To investigate the effect of sublaminar polyester tension bands on the biomechanics of the motion segments proximal to a long fusion construct.
Methods:
This was a human cadaveric biomechanical study. Pure moments of 4 Nm and 8 Nm were applied to the native spine and the instrumented spine, respectively (n = 8). The test conditions included native spine (T7-L2), fused (T10-L2), fused + bilateral tethers tensioned to 250 N at T9-T10 (tethers 250 N), fused + tethers tensioned to 350 N (tethers 350 N), fused (T11-L2) + tethers tensioned to 250 N at T9-T10 and 350 N at T10-T11 (2-level tethers), fused (T10-L2) + hand-tied suture loop through the spinous processes at T9-T10 (suture loop), and fused (T10-L2) with the T9-T10 interspinous and supraspinous ligaments cut (cut ISL/SSL).
Results:
The flexion range of motion (ROM) at T9-T10 of the fused spine, loaded at 8 Nm, increased to 162% of the native spine loaded at 4 Nm. The average flexion ROM at T9-T10 for tethers 250 N, tethers 350 N, 2-level tethers, suture loop, and cut ISL/SSL were 85% (P < 0.0001), 70% (P < 0.0001), 93% (P < 0.0001), 141% (P = 0.13), and 177% (P = 0.66) of the native spine at 4 Nm, respectively (P values vs. fused).
Conclusions:
Sublaminar polyester bands can modulate the biomechanical flexion ROM as a function of the band pretension and provide a more consistent and tunable technique than hand-tying a suture loop between the spinous processes.
Related Concept Videos
Tension Response at Adherens Junctions
α-Catenin as a Mechanosensory Protein
The α-catenin of adherens junctions is an allosteric protein with three VH (vinculin...
Band Theory
The energy difference between these bands is known as the band gap.
Conductor, Semiconductor,...
Transcription Attenuation in Prokaryotes
There are several different mechanisms used to attenuate transcription. In ribosome mediated...
Tension
Surface Tension, Capillary Action, and Viscosity
The various IMFs between identical molecules of a substance are examples of cohesive forces. The molecules within a liquid are surrounded by other molecules and are attracted equally in all directions by the cohesive forces within the liquid. However, the molecules on the surface of a liquid are attracted only by about one-half as many molecules. Because of the unbalanced molecular attractions on the surface molecules, liquids contract to form a shape that minimizes the number...
Types of Step-Growth Polymers: Polyesters
Polyesters are commonly prepared from terephthalic acid and ethylene glycol; the crude product is known as poly(ethylene terephthalate) or PET. However, polyesters are synthesized industrially by transesterification of dimethyl terephthalate with ethylene glycol at 150 °C. The two reactants and the polymer...