Validation and Cross-Reactivity Data for Fentanyl Analogs With the Immunalysis Fentanyl ELISA
Davide Guerrieri1, Fanny Kjellqvist2, Robert Kronstrand1,2
1Department of Medical and Health Science, Division of Drug Research, Linköping University, Linköping, Sweden.
Abstract:
Every year new fentanyl analog compounds, or fentanyls, appear on the drug scene. Development of immunoassays dedicated for screening individual molecules is challenging due to the short-lived presence of these compounds on the recreational drug market. Therefore, we investigated the detecting capabilities of the immunalysis fentanyl direct enzyme-linked immunosorbent assay (ELISA) kit against fentanyl in whole blood, and determined the cross-reactivity of nine fentanyl analogs (2-fluorofentanyl, acetylfentanyl, acrylfentanyl, carfentanil, cyclopropylfentanyl, tetrahydrofuranylfentanyl, furanylfentanyl, ocfentanil, valerylfentanyl) to confirm its validity for the general screening of fentanyls. Immunalysis ELISA assay was used to test whole blood samples fortified with fentanyl on a TECAN Freedom EVOlyzer platform, according to manufacturer specifications. The kit successfully was validated for fentanyl screening with a cutoff set at 0.5 ng/mL, and all tested analogs, with the exclusion of carfentanil, were detected. The lowest cross-reactivity with the kit was obtained with furanylfentanyl (20% ± 1, 95% confidence intervals (CI)) and 4-fluoroisobutyrfentanyl (25% ± 1, 95% CI), while the highest was recorded using acetylfentanyl (99% ± 11, 95% CI) and acrylfentanyl (94% ± 10, 95% CI). Post-mortem samples containing fentanyl, acrylfentanyl, cyclopropylfentanyl, THF-fentanyl and 4-fluoroisobutyrfentanyl were screened, and sensitivity and specificity of each analog were calculated. Positive screening results were generated by all post-mortem cases containing fentanyl (n = 14), acrylfentanyl (n = 11), cyclopropylfentanyl (n = 14), tetrahydrofuranylfentanyl (n = 13) and 4-fluoroisobutyrfentanyl (n = 10). Concentration of post-mortem fentanyl samples ranged from 0.5 ng/mL (cutoff) to 230 ng/mL, while the range for analogs was 3.4-36 ng/mL (cyclopentylfentanyl), 0.76-370 ng/mL (4-fluoroisobutyrfentanyl), 0.02-12 ng/mL (acrylfentanyl) and 2-26 ng/mL (tetrahydrofuranylfentanyl). The immunalysis fentanyl direct ELISA kit was successfully validated and showed significant cross-reactivity for all tested fentanyls, except carfentanil, making it a suitable technique for fentanyl and fentanyl analogs screening.
Related Concept Videos
Cross-reactivity
Data Validation
Key parameters for method validation include:
Data Validation
Nursing assessment guides are generally based on holistic models rather than medical...
Reliability and Validity
Crossing Over
The homologous pairs of sister chromosomes—one from the maternal and one from the paternal genome—then begin to align alongside each other lengthwise, matching corresponding DNA positions in a process...
Crossing Over