Higher HbA1c Measurement Quality Standards are Needed for Follow-Up and Diagnosis: Experience and Analyses from

Lutz Heinemann1, Patricia Kaiser2, Guido Freckmann3

  • 1Science Consulting in Diabetes GmbH, Neuss, Germany.

Hormone and Metabolic Research = Hormon- Und Stoffwechselforschung = Hormones Et Metabolisme
|September 15, 2018
PubMed

Insights

New quality controls for Hemoglobin A1c (HbA1c) testing are proposed to improve diabetes patient care. Narrower limits for internal and external quality assurance will enhance diagnostic accuracy and treatment decisions.

Area of Science:

  • Laboratory Medicine
  • Clinical Chemistry
  • Diabetology

Background:

  • Hemoglobin A1c (HbA1c) measurement is crucial for diabetes management and therapy decisions.
  • Current German guidelines (Rili-BAEK) allow wide deviations (±18% external, ±10% internal quality control), potentially masking imprecise results.
  • The introduction of unprocessed fresh EDTA blood for HbA1c testing necessitates stricter quality control.

Purpose of the Study:

  • To advocate for narrower acceptance limits for HbA1c internal (±3%) and external (±8%) quality controls in Germany.
  • To address the inadequacy of current broad deviation limits in ensuring accurate HbA1c measurements for patient care.
  • To highlight the need for improved quality assurance, especially for point-of-care testing (POCT) devices.

Main Methods:

  • Analysis of the current Rili-BAEK guidelines for HbA1c external quality assessment (EQA).
  • Evaluation of the impact of new EQA materials and unprocessed fresh EDTA blood on measurement accuracy.
  • Review of the necessity for stricter quality control criteria in laboratory medicine.

Main Results:

  • Existing broad deviation limits in EQA are insufficient for reliable HbA1c patient care.
  • New, narrower acceptance criteria (±3% internal, ±8% external) are proposed for German EQA schemes.
  • Stricter controls are essential to identify imprecisely measured HbA1c results and incorrectly calibrated devices.

Conclusions:

  • Implementing tighter quality control limits for HbA1c measurements is vital for accurate diabetes diagnosis and treatment.
  • Improved HbA1c measurement quality will prevent misdiagnosis and inappropriate patient management.
  • Harmonizing quality standards will enhance patient safety and therapeutic efficacy in diabetes care.

Related Concept Videos

Measurement: Standard Units03:38

Measurement: Standard Units

Every measurement provides three kinds of information: the size or magnitude of the measurement (a number), a standard of comparison for the measurement (a unit), and an indication of the uncertainty of the measurement. While the number and unit are explicitly represented when a quantity is written, the uncertainty is an aspect of the errors in the measurement results.
80.4K
Units and Standards of Measurement01:10

Units and Standards of Measurement

A physical quantity is defined either by specifying its measurement method or by stating how it is calculated from other measurements. For example, consider a metallic cube. We might define its mass and dimensions by specifying methods for measuring them, such as using a weighing machine and a meter scale. Then, we could define the volume by stating that it is the cube of its side, and we could calculate the density as the mass divided by the volume.
Measurements of physical quantities are...
45.3K
The Stanford Prison Experiment03:20

The Stanford Prison Experiment

The famous and controversial Stanford Prison Experiment, conducted by social psychologist Philip Zimbardo and his colleagues at Stanford University, demonstrated the power of social roles, social norms, and scripts.
24.7K
Standard Electrode Potentials03:02

Standard Electrode Potentials

On comparing the reactivity of silver and lead, it is observed that the two ionic species, Ag+ (aq) and Pb2+ (aq), show a difference in their redox reactivity towards copper: the silver ion undergoes spontaneous reduction, while the lead ion does not. This relative redox activity can be easily quantified in electrochemical cells by a property called cell potential. This property is commonly known as cell voltage in electrochemistry, and it is a measure of the energy which accompanies the charge...
50.4K
Standard Enthalpy of Formation02:37

Standard Enthalpy of Formation

Enthalpy changes are typically tabulated for reactions in which both the reactants and products are at the same conditions. A standard state is a commonly accepted set of conditions used as a reference point for the determination of properties under other different conditions. For chemists, the IUPAC standard state refers to materials under a pressure of 1 bar and solutions at 1 M and does not specify a temperature. Many thermochemical tables list values with a standard state of 1 atm. Because...
49.3K
Nursing Diagnosis01:22

Nursing Diagnosis

Following assessment, a nursing diagnosis is the next step in the nursing process. It begins after the nurse has collected and recorded the patient data. The purpose of diagnosing is to identify how the client responds to actual or potential health processes, identify factors that bestow or that cause health problems, the etiologies, and identify resources or strengths the individual, group, or community can draw on to prevent or resolve problems.
The nursing diagnosis focuses on evidence-based...
4.1K