Impact of Connector Placement and Design on Bending Stiffness of Spinal Constructs
Jakub Godzik1, Nick Hool2, Jonathan F Dalton3
1Department of Neurosurgery, Barrow Neurological Institute, St. Joseph's Hospital and Medical Center, Phoenix, Arizona.
Objective:
To evaluate the stability of multiple rod-connector construct designs using a mechanical 4-point bending testing frame.
Methods:
A mechanical study was used to evaluate the bending stiffness of 3 connectors across 12 different configurations of rod-connector-rod constructs. Stability was evaluated in flexion-extension and lateral bending. Combinations of rods having 1 of 3 diameters (4.0 mm, 5.5 mm, and 6.0 mm) connected by 1 of 3 connector types (parallel open, snap-on, and hinged) were compared. Configurations with single connectors and with double connectors with variable spacing were also compared to simulate revision surgery conditions.
Results:
Constructs consisting of 4.0-mm rods connected to 4.0-mm rods were significantly less stiff as the total number of connectors used in a series exceeded 2. When single-connector configurations were compared, parallel open rod connectors demonstrated greater stiffness in flexion-extension than hinged open connectors, whereas hinged open connectors demonstrated greater stiffness in lateral bending. Using double connectors increased stiffness of 4.0- to 4.0-mm rod configurations in flexion-extension and lateral bending, 4.0- to 6.0-mm rod configurations in flexion-extension, and 5.5- to 6.0-mm rod configurations in lateral bending. Spacing the double connectors significantly improved lateral bending stiffness of 4.0- to 4.0-mm and 5.5- to 6.0-mm rod configurations.
Conclusions:
Our data indicate that the design, number, and placement of rod connectors have a significant impact on the bending stiffness of a surgical construct. Such mechanical data may influence construct design in primary and revision surgeries of the cervical spine and cervicothoracic junction.
Related Concept Videos
Design of Prismatic Beams for Bending
Design Example: Dimensioning of Concrete Masonry Construction
The site engineer has laid out a plan for the storeroom with external dimensions of twelve feet in length and...
Bending
In pure bending, the bending stress in a beam is calculated based on the bending moment and...
Symmetric Member in Bending
Unsymmetric Bending
Bending of Members Made of Several Materials
Hooke's Law determines stress in each material, stating that stress is proportional to strain but varies due to each material's...