Related Experiment Video
Updated: Feb 5, 2026

Transcatheter Pulmonary Valve Replacement from Autologous Pericardium with a Self-Expandable Nitinol Stent in an Adult Sheep Model
Published on: June 8, 2022
Optimal positioning of self-expanding valves before deployment decreases paravalvular regurgitation following
Razvan T Dadu1, Jiaqiong Xu1, Hasan Rehman1
1Department of Cardiology, Houston Methodist DeBakey Heart and Vascular Center, Houston, Texas.
Objectives:
To evaluate the association between measurements performed during Medtronic CoreValve (MCV) deployment and paravalvular leak (PVL).
Background:
The MCV can be recaptured and repositioned, allowing the TAVR operator to implant at a more favorable position. The association between angiographic measurements of MCV position while the valve is recapturable and PVL post deployment has not been investigated.
Methods:
493 patients undergoing TAVR with MCV (January 2011-July 2017) were included. PVL was defined as intraprocedural aortic regurgitation that was judged clinically to require balloon postdilation. Depth of the valve at the left coronary cusp (LCC) and noncoronary cusp (NCC) were measured when the valve was 80% deployed. An optimal cutoff value for the ratio LCC/NCC for PVL was identified in 40 patients. Using this cutoff value, the association between LCC/NCC and PVL was then validated in 453 patients.
Results:
The median LCC/NCC was 1.51 (interquartile range 1.06-1.89).The optimal cutoff value for LCC/NCC was 1.48 (93% sensitivity, 77% specificity, AUC0.85). In the validation group 112 (24.7%) patients had PVL. For LCC/NCC ≥ 1.48, the incidence of PVL was lower compared to LCC/NCC < 1.48 (9.58% vs. 41.78%, P < 0.0001). LCC/NCC of 1.48 had a sensitivity of 79.5% and specificity of 63.6% for PVL (AUC0.72). In a multivariate model, LCC/NCC < 1.48 independently predicted PVL (OR = 6.67, 95% CI 3.96-11.23, P < 0.0001).
Conclusion:
Positioning the MCV such that the LCC/NCC is ≥1.48 may result in less PVL.
Related Concept Videos
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Aortic Regurgitation I: Introduction
Mitral Valve Prolapse I: Introduction
Aortic Regurgitation III: Medical Management
Aortic Regurgitation IV: Nursing Management
Mitral Valve Prolapse II: Assessment and Management