A Dual-Mode Imaging Catheter for Intravascular Ultrasound Application
Insights
This study introduces a novel dual-mode imaging catheter combining ultrasound and flow measurement. This innovative device enables simultaneous assessment of arterial morphology and blood flow, aiding in cardiovascular disease diagnosis.
Area of Science:
- Cardiovascular imaging
- Medical device engineering
- Biomedical ultrasound
Background:
- Arterial morphology and blood flow are crucial for cardiovascular disease evaluation.
- Current intravascular imaging methods lack functional flow data.
- Existing flow assessment tools do not provide cross-sectional arterial images.
Purpose of the Study:
- To design and fabricate a novel dual-mode imaging catheter.
- To integrate forward-looking and side-looking ultrasonic transducers in a single catheter.
- To enable simultaneous acquisition of arterial morphology and functional flow parameters.
Main Methods:
- Fabrication of a unique dual-mode catheter housing both forward-looking and side-looking ultrasonic transducers.
- Inclusion of components for precise blood flow measurement.
- Testing of the catheter's capability to acquire multiple arterial parameters.
Main Results:
- The dual-mode catheter successfully provides both morphological and functional flow data.
- Multiple arterial parameters can be acquired during a single, one-time procedure.
- The device demonstrates potential for comprehensive arterial assessment.
Conclusions:
- The developed dual-mode imaging catheter offers a unified approach for evaluating arterial health.
- This novel catheter facilitates simultaneous morphological and functional assessment of arteries.
- The technology holds promise for improved functional diagnosis of atherosclerotic arteries.
Abstract:
Both the morphological anatomy and functional parameters such as flow speed of the artery provide valuable information for the evaluation of cardiovascular diseases. Direct measurement of the arterial wall can be achieved by intravascular optical/ultrasound imaging methods, and however, no functional data are acquired with these methods. Fractional flow reserve and Doppler wire have been used to assess the blood flow information, but do not provide cross-sectional images of the artery. This paper is the first to design and fabricate a dual-mode imaging catheter that contains a forward-looking ultrasonic transducer and a side-looking ultrasonic transducer together in one catheter. This dual-mode catheter not only provides morphological information about the artery, but also a precise measurement of functional flow. The data indicate that the proposed catheter can be used to acquire multiple parameters of the artery with a one-time procedure. This novel one-catheter approach could be used for the functional diagnosis of atherosclerotic arteries.
Related Concept Videos
What is a Mode?
There can be more than one mode in a data set if multiple values have the same highest frequency. For instance, suppose that the Statistics exam scores of 20 students are: 50; 53; 59; 59; 63; 63; 72; 72; 72; 72; 72; 76; 78; 81; 83; 84; 84; 84; 90; 93. Here, the mode is 72, as it occurs most frequently, five times.
A data set with two modes is called bimodal. For example,...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Modes of Standing Waves - I
Modes of Standing Waves: II
For a tube open at one end and closed at the other filled with air, the modes are such that there is always an antinode at the open end and a node at the closed end....
MOSFET: Enhancement Mode
In their basic form, enhancement-mode MOSFETs are typically non-conductive when the gate-source voltage (Vgs) is zero. This default 'off' state means no...
Modes of Operations of BJT
Active Mode: The most common mode for amplification, the active mode features a forward-biased emitter-base junction and a reverse-biased base-collector junction. This setup enables electrons to be injected from the emitter to the base while blocking the majority carriers at the collector. The...