Related Experiment Video
Updated: Feb 5, 2026

An Experimental Platform to Study the Closed-loop Performance of Brain-machine Interfaces
Published on: March 10, 2011
Performance of closed-loop resuscitation of haemorrhagic shock with fluid alone or in combination with
Nicolas Libert1,2, Guillaume Chenegros3, Anatole Harrois1,4
1Laboratoire d'Etude de la Microcirculation, UMR 942, Université Paris 7, Hôpitaux Saint Louis Lariboisière, Assistance Publique-Hôpitaux de Paris, Paris, France.
Background:
Closed-loop resuscitation can improve personalization of care, decrease workload and bring expert knowledge in isolated areas. We have developed a new device to control the administration of fluid or simultaneous co-administration of fluid and norepinephrine using arterial pressure.
Method:
We evaluated the performance of our prototype in a rodent model of haemorrhagic shock. After haemorrhagic shock, rats were randomized to five experimental groups: three were resuscitated with fluid and two with co-administration of fluid and norepinephrine. Among groups resuscitated with fluid, one was resuscitated by a physician and two were resuscitated according to two different closed-loop algorithms. Among groups resuscitated with fluid and norepinephrine, one was resuscitated by a physician and the other one by the closed-loop device. The precision of arterial pressure during the resuscitation period was assessed using rising time, time passed in the target area and performance error calculations.
Results:
Groups resuscitated with fluid had similar performances and passed as much time in the target area of 80-90 mmHg as the manual group [manual: 76.8% (67.9-78.2), closed-loop: 64.6% (45.7-72.9) and 80.9% (59.1-85.3)]. Rats resuscitated with fluid and norepinephrine using closed-loop passed similar time in target area than manual group [closed-loop: 74.4% (58.4-84.5) vs. manual: 60.1% (46.1-72.4)] but had shorter rising time to reach target area [160 s (106-187) vs. 434 s (254-1081)] than those resuscitated by a physician. Rats resuscitated with co-administration of fluid and norepinephrine required less fluid and had less hemodilution than rats resuscitated with fluid alone. Lactate decrease was similar between groups resuscitated with fluid alone and fluid with norepinephrine.
Conclusions:
This study assessed extensively the performances of several algorithms for closed-loop resuscitation of haemorrhagic shock with fluid alone and with co-administration of fluid and norepinephrine. The performance of the closed-loop algorithms tested was similar to physician-guided treatment with considerable saving of work for the caregiver. Arterial pressure closed-loop guided algorithms can be extended to combined administration of fluid and norepinephrine.
Related Concept Videos
Open and closed-loop control systems
An open-loop control system operates without feedback from the output. It consists of two primary elements: the controller and the controlled process. The controller receives an input signal...
Cardiopulmonary Resuscitation I: Adult
Shock Waves
When the source's speed approaches the speed of sound, constructive interference between successive wavefronts emitted by the source occurs immediately behind it. Initially, scientists believed that this constructive interference would result in such high...
Cardiopulmonary Resuscitation III: AED Use
Cardiopulmonary Resuscitation IV: Pharmacological Management
Feedback Loops