Related Experiment Video
Updated: Feb 5, 2026

Designing CAD/CAM Surgical Guides for Maxillary Reconstruction Using an In-house Approach
Published on: August 24, 2018
Evaluation of Vertical Misfit of CAD/CAM Implant-Supported Titanium and Zirconia Frameworks
Purpose:
This study compared the vertical misfit of fixed implant-supported titanium (Ti6Al4V) and yttria-stabilized tetragonal zirconia frameworks milled using the computer-aided design/computer-aided manufacturing (CAD/CAM) Zirkonzahn system.
Materials And Methods:
An aluminum mandibular master cast was used, and four analogs of the Regular CrossFit implant for straight multibase titanium alloy abutments (Institut Straumann) were fixed with self-curing acrylic resin in mandibular canine and second premolar tooth locations with the aid of a parallelometer. Scan abutments were placed over the four straight multibase Regular CrossFit abutments to determine their correct three-dimensional (3D) positioning in the virtual model. Implant frameworks were designed virtually, and five frameworks of each material (titanium alloy or yttria-stabilized tetragonal zirconia) were milled using the CAD/CAM system. Scanning electron microscope (SEM) images, with magnifications of 50×, 250×, and 1,000×, were obtained from three points on the buccal surface, three points on the lingual surface, and three points on the mesial surface of each abutment. The measurements were performed in the center of each of the thirds, thus defining points one, two, and three. The degree of vertical misfit at the framework-abutment interface was determined using ImageJ software and analyzed using the Student t test (P = .05).
Results:
The mean misalignment value was 6.011 ± 0.750 μm for the titanium group and 9.055 ± 3.692 μm for the zirconia group.
Conclusion:
Within the limitations of this study, there was no significant difference in vertical misfit between the titanium and zirconia implant frameworks produced using the CAD/CAM system. Regardless of the material used, the CAD/CAM frameworks achieved an adequate vertical fit.
More Related Videos
07:05In Vitro Evaluation of The Effects Of Er,Cr:YSGG and Diode Lasers Used on Titanium Cylinder
Published on: June 6, 2025
08:19Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
Related Concept Videos
C4 Pathway and CAM
C4 Pathway
The C4 pathway is used by plants such as...
Vertical Curve: Problem Solving
Self-Help Support Groups
Accessibility and Cost-Effectiveness
One of the primary strengths of self-help...
Introduction to Vertical Curves
Support Reactions
The purpose of the supports is to prevent the translational motion of the system by applying an equal and opposite force and to prevent the system's rotation by applying...
Elevation of Intermediate Points on Vertical Curves