3D task-transfer function representation of the signal transfer properties of low-contrast lesions in FBP- and
Marthony Robins1, Justin Solomon2, Taylor Richards1
1Carl E. Ravin Advanced Imaging Laboratories, Department of Radiology, Medical Physics Graduate Program, Duke University Medical Center, Durham, NC, 27705, USA.
Purpose:
The purpose of this study was to investigate how accurately the task-transfer function (TTF) models the signal transfer properties of low-contrast features in a non-linear commercial CT system.
Methods:
A cylindrical phantom containing 24 anthropomorphic "physical" lesions was 3D printed. Lesions had two sizes (523, 2145 mm3 ), and two nominal radio-densities (80 and 100 HU at 120 kV). CT images were acquired on a commercial CT system (Siemens Flash scanner) at four dose levels (CTDIvol , 32 cm phantom:1.5, 3.0, 6.0, 22.0 mGy) and reconstructed using FBP and IR kernels (B31f, B45f, I31f\2, I44f\2). Low-contrast rod inserts (in-plane) and a slanted edge (z-direction) were used to estimate 3D-TTFs. CAD versions of lesions were blurred by the 3D-TTFs, virtually superimposed into corresponding phantom images, and compared to the physical lesions in terms of (a) a 4AFC visual assessment, (b) edge gradient, (c) size, and (d) shape similarity. Assessments 2 and 3 were based on an equivalence criterion to determine if the natural variability in the physical lesions was greater or equal to the difference between physical and simulated. Shape similarity was quantified via Sorensen-Dice coefficient (SDC). Comparisons were done for each lesion and for all imaging conditions.
Results:
The readers detected simulated lesions at a rate of 37.9 ± 3.1% (25% implies random guessing). Lesion edge blur and volume differences were on average less than physical lesions' natural variability . The SDC (average ± SD) was 0.80 ± 0.13 (max of 1 possible).
Conclusions:
The visual appearance, edge blur, size, and shape of simulated lesions were similar to the physical lesions, which suggests 3D-TTF models the low-contrast signal transfer properties of this non-linear CT system reasonably well.
Related Concept Videos
Transfer Function to State Space
In an RLC...
State Space to Transfer Function
The transformation process begins with the state-space representation, characterized by the state equation and the output equation. These equations are typically represented as:
Transfer function and Bode Plots-II
Transfer function and Bode Plots-I
Transfer Function in Control Systems
To derive the transfer function, consider a general nth-order linear time-invariant...
Reconstruction of Signal using Interpolation