Comparing the Effects of Additives on Protein Analysis Between Desorption Electrospray (DESI) and Electrospray
Elahe Honarvar1, Andre R Venter2
1Department of Chemistry, Western Michigan University, Kalamazoo, MI, 49008-5413, USA.
Abstract:
It is frequently said that DESI-MS follows a similar ionization mechanism as ESI because of similarities usually observed in their respective mass spectra. However, practical use of DESI-MS for protein analysis is limited to proteins with lower molecular weights (< 25 kDa) due to a mass-dependent loss in signal intensity. Here we investigated commonly used volatile acids and their ammonium salt buffers for DESI-MS analysis of protein. We noticed that, surprisingly, some additives influence the analysis differently in DESI compared to ESI. Improved signal intensities with both DESI and ESI were obtained when acetic and formic acid were added into aqueous methanol spray solvents with both DESI and ESI. On the other hand, while with ESI the addition of ammonium salts into spray solutions strongly reduced both signal and S/N, with DESI signal intensities and S/N were improved dramatically. Ammonium bicarbonate when used with DESI reduced the total amount of adduction and delivered excellent signal-to-noise ratios with high intensity; however, it also denatures protein. When native state protein mass spectra are preferred, ammonium acetate would also deliver reasonable adduct removal and improved S/N. The amount of total adduction of individual adducting species and of all species could not be correlated with differences in either solutions pH values or with proton affinities of the anions. An obvious difference between DESI and ESI mass spectrometry is the effects of protein solubility during droplet pickup (desorption), but differences in the sizes, velocities, and composition of ionizing droplets were also discussed as important factors. Graphical Abstract ᅟ.
More Related Videos
Related Concept Videos
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Matrix-Assisted Laser Desorption Ionization (MALDI)
Ionization Energy
Comparing the Survival Analysis of Two or More Groups
Conjugate Addition (1,4-Addition) vs Direct Addition (1,2-Addition)
Conjugate addition results in a thermodynamically stable product. The reaction retains the stronger C=O bond at the expense of the weaker C=C π bond. The process is slow as the β carbon is less electrophilic than the carbonyl carbon.
Direct addition products are...
Biological Effects of Radiation