Comparison of 7 Different Sensors for Detecting Low Respiratory Rates Using a Single Breath Detection Algorithm in
Sean Ermer1, Lara Brewer, Joe Orr
1From the Department of Anesthesia, University Of Utah, Salt Lake City, Utah.
Background:
Numerous technologies are used to monitor respiratory rates in nonintubated patients. No technology has emerged as the standard. The primary aim of this study was to assess the limits of agreement between a reference sensor signal (respiratory inductance plethysmography bands) and 7 alternative sensor signals (nasal capnometer, nasal pressure transducer, oronasal thermistor, abdominal accelerometer, transpulmonary electrical impedance, peritracheal microphone, and photoplethysmography) for measuring low respiratory rates in sedated, nonintubated, supine volunteers. A unified approach based on a single breath detection algorithm was applied to each sensor to facilitate comparison. We hypothesized that all of the sensor signals would allow detection of low (<10 breaths per minute) respiratory rates to within ±2 breaths per minute of the reference sensor signal.
Methods:
Volunteers received remifentanil and propofol infusions at selected target concentration pairs to induce depression of ventilation. Signals from each sensor were analyzed by an identical threshold-based detection algorithm to compute the breathing rate. Bland-Altman limits of agreement and error rate analyses were used to characterize the performance of each sensor compared to the reference sensor.
Results:
The analysis of the accelerometer and capnometer signals, using Bland-Altman and error rate analyses, showed the highest breath rate agreement (1.96 × standard deviation) of the 7 sensors with -2.1 to 2.2 and -2.5 to 2.7 breaths per minute, respectively. All other signals exhibited wider limits of agreement, with impedance being the widest at -7.8 to 7.4 breaths per minute. For the abdomen accelerometer, 95% of Bland-Altman data points were within ±2 breaths per minute. For the capnometer, 96% of data points were within ±2 breaths per minute. Nasal pressure, thermistor, and microphone all had >80% of data points within ±2 breaths per minute. Impedance and photoplethysmograph signals had 58% and 64%, respectively.
Conclusions:
A unified approach can be applied to a variety of sensor signals to estimate respiratory rates in spontaneously breathing, nonintubated, sedated volunteers. However, detecting clinically relevant low respiratory rates (<6 breaths per minute) is a technical challenge. By our analysis, no single sensor was able to detect slow respiratory rates with adequate precision (<±2 breaths per minute of the reference signal). Of the sensors evaluated, capnometers and abdominal accelerometers may be the most reliable sensors for identifying hypopnea and central apnea.
More Related Videos
05:16Flying Insect Detection and Classification with Inexpensive Sensors
Published on: October 15, 2014
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
Related Concept Videos
Breathing
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Sedatives and Hypnotics: Overview
Sedative-hypnotics are categorized into barbiturates, benzodiazepines (BZDs), and non-benzodiazepines or Z-drugs. These drugs work by suppressing central nervous system activity, and this suppression is dose-dependent. Older sedative medications, like barbiturates, follow a linear curve in...
Detection of Gross Error: The Q Test
Detection of Black Holes
Their closest cousins are neutron stars, which are composed almost entirely of neutrons packed against each other, making them extremely dense. A neutron star has the same mass as the Sun but its diameter is only a few kilometers. Therefore, the escape velocity from their surface is close to the speed of light.
Not until the 1960s, when the first neutron...
Sedatives and Hypnotics Drugs: Barbiturates