Biomechanical Comparison of C1 Lateral Mass-C2 Short Pedicle Screw-C3 Lateral Mass Screw-Rod Construct Versus
Young Lu1, Yu-Po Lee1, Nitin N Bhatia1
1Department of Orthopedics, University of California, Irvine, CA.
Study Design:
A biomechanical in vitro study using human cadaveric spines.
Objective:
The aim of this study was to compare atlantoaxial stability and stiffness of a C1 lateral mass - C2 short pedicle - C3 lateral mass screw-rod construct versus C1 lateral mass - C2 pedicle screw-rod construct.
Summary Of Background Data:
The C1 lateral mass - C2 pedicle screw-rod construct provides excellent atlantoaxial fixation, but C2 pedicle screw placement is associated with risk of vertebral artery injury. The use of shorter C2 pedicle screws may mitigate the risk of vascular injury but may result in reduced C1-C2 stabilization. Extending C1 lateral mass - C2 short pedicle screw-rod construct with C3 lateral mass screws may mitigate the risk of vascular injury without compromising C1-C2 fixation.
Methods:
Seven cervical spines were tested with internal control experimental design in the following sequence: intact state, and following creation of type II odontoid fracture, the specimen was instrumented with C1 lateral mass - C2 pedicle screw fixation (C2PED), C1 lateral mass - C2 short pedicle screw fixation (C2SPED), and C1 lateral mass - C2 short pedicle - C3 lateral mass screw fixation (C2SPED-C3LM). For each condition, the angular stiffness and range of motion (ROM) with 1.5-Nm load in flexion/extension (FE), lateral bending (LB), and right/left axial rotation (RAR/LAR) were quantified.
Results:
Instrumented conditions demonstrated significantly lower C1-C2 angular ROM and greater stiffness than the intact state. Compared with C2PED, C2SPED-C3LM demonstrated significantly lower C1-C2 ROM during FE and LB, significantly greater C1-C2 stiffness in flexion and right/left LB, similar C1-C2 ROM and stiffness in RAR/LAR, and similar stiffness in extension. C2SPED-C3LM demonstrated significantly greater atlantoaxial stabilization in the sagittal and coronal planes than C2PED construct.
Conclusion:
Compared with C2PED, C2SPED-C3LM may be a suitable alternative surgical strategy for atlantoaxial instability that provides superior atlantoaxial fixation.
Level Of Evidence:
N/A.
Related Concept Videos
Screw: Problem Solving
To evaluate the...
Self-Locking Screw
A square-threaded screw jack carrying a load is considered self-locking if the screw retains its position even after the moment applied to it is removed.
Frictional Forces on Screws
A jack with a square-threaded screw is a mechanical device used to lift heavy loads by applying a force at its handle. When the force is applied, the screw turns, raising the load. The screw can...
Lateralization
Atomic Mass
Molar Mass