Reduced Procedure Time and Variability with Active Esophageal Cooling During Radiofrequency Ablation for Atrial Fibrillation
Published on: August 25, 2022
Prospective Multicenter Experience With Cooled Radiofrequency Ablation Using High Impedance Irrigant to Target Deep
Duy T Nguyen1, Wendy S Tzou1, Amneet Sandhu1
1Section of Cardiac Electrophysiology, University of Colorado, Aurora, Colorado.
Objectives:
This study sought to evaluate the efficacy and safety of using half-normal saline (HNS) as the cooling radiofrequency ablation (RFA) irrigant among patients who had failed prior, standard RFA.
Background:
Effective control of ventricular arrhythmias that arise from mid-myocardium may be refractory to standard RFA. Recent data suggest that delivering fluid with decreased ionic concentration during open-irrigated RFA can produce deeper RFA lesions.
Methods:
A 12-center prospective analysis was performed of all ablations using HNS for the treatment of ventricular arrhythmias (premature ventricular complex [PVC]/ventricular tachycardia [VT]) refractory to standard ablation with normal saline irrigant.
Results:
HNS RFA was used clinically to target 94 PVC/VTs refractory to standard ablation. Acute success was achieved in 78 of 94 (83%), with longer-term success occurring in 78 subjects after a mean follow-up of 6.1 ± 6.7 months (range, 3.0 to 25.2 months). Steam pops were observed among 12 (12.6%) patients. There were no significant changes in electrolytes measured before and after the use of HNS, and there were no complications related to HNS use.
Conclusions:
The use of HNS instead of normal saline irrigant during high-power delivery targeting deep myocardial substrate is safe and effective. PVC/VT sources previously unaffected by standard ablation may be successfully ablated with improved efficiency of radiofrequency delivery using HNS.
Related Concept Videos
Heating and Cooling Curves
For instance, the addition of heat raises the temperature of a solid; the amount of heat absorbed depends on the heat capacity of the solid (q = mcsolidΔT). According to thermochemistry, the relation between the amount of heat absorbed or released by a substance, q, and its...
Design Example: Design of an Irrigation Channel
Impedance Combination
Impedances and Admittance
Impedance is a measure of resistance to sinusoidal current flow in an AC circuit. Unlike their behavior in DC circuits, where inductors appear as short circuits and capacitors as open circuits, the behavior of these components in AC...
What is an Experiment?
Series Impedances: Three-Phase Line
Using Kirchhoff's laws, an integro-differential equation for the network is derived. This equation accounts for unbalanced phase currents, which may induce return currents through neutral wires and the earth, seeking the least impedance path. Earth return conductors can replace the...