A Novel Intramedullary Locked Fixation Device for Treatment of Clavicle Shaft Fractures
Paul R King1, Carl J Basamania2, Robert P Lamberts1
1Division of Orthopaedic Surgery, Department of Surgical Sciences, Stellenbosch University, Tygerberg Campus, Tygerberg, South Africa.
Introduction:
An intramedullary nail (the Sonoma CRx) with a unique flexible anatomical design and locking system is a viable alternative for treatment of displaced and shortened clavicle shaft fractures.
Step 1 Positioning:
Place the patient in the beach-chair position and drape the arm free so that you can obtain an exaggerated anteroposterior and an axial fluoroscopic view of the clavicle by manipulating the position of the arm.
Step 2 Skin Incision:
The surgical technique necessitates opening the fracture site in order to implant the device.
Step 3 Preparing The Medial Fragment:
The preparation of the medial fragment is of utmost importance as the intramedullary Sonoma CRx nail should be placed as far as possible into the medial fragment of the clavicle, to provide the highest level of stability and the lowest risk of device failure.
Step 4 Preparing The Lateral Fragment:
Preparation of the lateral fragment is more challenging than preparation of the medial fragment because of its relative immobility and shorter medullary canal.
Step 5 Placement Of The Intramedullary Nail:
Use the longest possible intramedullary nail that the clavicle can accommodate and insert the nail as far medially as possible into the medullary canal of the clavicle bone.
Step 6 Activation Of The Locking Device:
Lock the device at its medial end with distally deployed grippers and laterally with a locking screw placed through a jig.
Step 7 Management Of Comminuted Fracture Fragments:
Comminuted fractures are not a contraindication to the use of the intramedullary device as they can be reduced by the nail and secured by using cerclage sutures.
Step 8 Postoperative Management:
Protect the affected shoulder in a shoulder immobilizer for six weeks.
Results:
In our original study, forty-seven consecutive patients with a displaced and shortened clavicle shaft fracture were treated with open reduction and internal fixation using the Sonoma CRx device.
More Related Videos
07:41An Intramedullary Locking Nail for Standardized Fixation of Femur Osteotomies to Analyze Normal and Defective Bone Healing in Mice
Published on: November 13, 2016
15:11Surgical Fixation of Sternal Fractures: Preoperative Planning and a Safe Surgical Technique Using Locked Titanium Plates and Depth Limited Drilling
Published on: January 5, 2015
Related Concept Videos
Stresses in a Shaft
Applying equilibrium conditions to the QR segment establishes that the internal shearing forces within the...
Self-Locking Screw
A square-threaded screw jack carrying a load is considered self-locking if the screw retains its position even after the moment applied to it is removed.
Deformation in a Circular Shaft
Transmission Shafts: Problem Solving
Next, use bending moment diagrams for the shaft to...
Design of Transmission Shafts
Thin-Walled Hollow Shafts