Performance analysis of automated evaluation of antinuclear antibody indirect immunofluorescent tests in a routine
Joyce J B C van Beers1, Melanie Hahn2, Johanna Fraune2
1Central Diagnostic Laboratory, Maastricht University Medical Center, P. Debyelaan 25, 6229 HX, Maastricht, The Netherlands.
Purpose:
Indirect immunofluorescence (IIF) on the human epithelial cell-line HEp-2 (or derivatives) serves as the gold standard in antinuclear antibody (ANA) screening. IIF, and its evaluation, is a labor-intensive method, making ANA testing a major challenge for present clinical laboratories. Nowadays, several automated ANA pattern recognition systems are on the market. In the current study, the EUROPattern Suite is evaluated for its use in daily practice in a routine setting.
Methods:
A total of 1033 consecutive routine samples was used to screen for ANA. Results (positive/negative ANA screening, pattern identification and titer) were compared between software-generated results (EUROPattern) and visual interpretation (observer) of automatically acquired digital images.
Results:
Considering the visual interpretation as reference, a relative sensitivity of 99.3% and a relative specificity of 88.9% were obtained for negative and positive discrimination by the software (EPa). A good agreement between visual and software-based interpretation was observed with respect to pattern recognition (mean kappa: for 7 patterns: 0.7). Interestingly, EPa software distinguished more patterns per positive sample than the observer (on average 1.5 and 1.2, respectively). Finally, a concordance of 99.3% was observed within the range of 1 titer step difference between EPa and observer.
Conclusions:
The ANA IIF results reported by the EPa software are in very good agreement with the results reported by the observer with respect to being negative/positive, pattern recognition and titer, making automated ANA IIF evaluation an objective and time-efficient tool for routine testing.
Related Concept Videos
Indirect Motor Pathways
The vestibulospinal tract originates in the vestibular nuclei of the brainstem. The vestibular system detects changes in...
Performing a Simple Data Analysis using MS-Excel Function
SUM: This function calculates the total sum of a range of values. It's the foundation for aggregating data, essential for determining overall trends and totals in datasets.
AVERAGE: It computes the mean value of a given set of numbers, providing a quick insight into the central...
EDTA: Indirect and Alkalimetric Titration
Antibody Structure
Antibodies, also known as immunoglobulins (Ig), are essential players of the adaptive immune system. These antigen-binding proteins are produced by B cells and make up 20 percent of the total blood plasma by weight. In mammals, antibodies fall into five different classes, which each elicits a different biological response upon antigen binding.
The Y-Shaped Structure of Antibodies Consists of Four Polypeptide Chains
Antibodies consist of four polypeptide chains: two identical heavy...
Self-Evaluation: Self-Enhancement and Self-Verification
Setting Time of Cement


![An Automated Radiosynthesis of [68Ga]Ga-FAPI-46 for Routine Clinical Use](/_next/image?url=https%3A%2F%2Fcloudfront.jove.com%2FCDNSource%2Fteasers%2F66708.jpg&w=3840&q=50)