Multicenter Study of Pipeline Flex for Intracranial Aneurysms
Leonardo B C Brasiliense1, Pedro Aguilar-Salinas2, Demetrius K Lopes3
1Division of Neurosurgery, University of Arizona, Tucson, Arizona.
The Pipeline Expandable Intracranial Aneurysm Flow Diverter (PED Flex) demonstrated low rates of neurological complications and mortality in a multicenter study. This device facilitates deployment, reducing technical events for treating brain aneurysms.
Area of Science:
- Neuroendovascular surgery
- Cerebrovascular disease management
- Medical device technology
Background:
- The Pipeline Flex (PED Flex) device was engineered to improve upon its predecessor, aiming for easier deployment and navigation.
- These design enhancements are intended to decrease the incidence of technical complications during neurovascular procedures.
Purpose of the Study:
- To evaluate the 30-day neurological morbidity and mortality rates associated with the PED Flex device.
- To assess the safety and efficacy of the PED Flex in treating intracranial aneurysms.
Main Methods:
- A retrospective analysis of 205 patients with 223 intracranial aneurysms treated across 9 neurovascular centers.
- Data collected included patient and aneurysm characteristics, periprocedural events, and clinical/angiographic outcomes.
- Multivariate logistic regression was used to identify predictors of unfavorable clinical outcomes (modified Rankin Scale > 2).
Main Results:
- The 30-day neurological morbidity and mortality rates were low at 1.9% and 0.5%, respectively.
- Intraprocedural events without neurological compromise occurred in 6.8% of patients.
- Favorable clinical outcomes (mRS 0-2) were achieved in 94.5% of patients at 30 days.
Conclusions:
- The PED Flex device is associated with low 30-day neurological morbidity and mortality rates for intracranial aneurysm treatment.
- Technical events related to device deployment were infrequent, likely due to design modifications.
- The study supports the safety and effectiveness of the PED Flex in a multicenter setting.