Automation of the Micronucleus Assay Using Imaging Flow Cytometry and Artificial Intelligence
Published on: January 27, 2023
Inter-laboratory validation of a harmonized PNH flow cytometry assay
Daniel Payne1, Ulrika Johansson2, David Bloxham3
1University Hospitals of Leicester, Leicester Royal Infirmary, Infirmary Square, Leicester, Leicestershire, LE1 5WW, United Kingdom.
A new network-validated assay for paroxysmal nocturnal hemoglobinuria (PNH) testing ensures reliable results across laboratories. This robust PNH assay improves precision and accuracy for diagnosing PNH.
Area of Science:
- Clinical diagnostics
- Hematology
- Flow cytometry
Background:
- A network of 11 laboratories aimed to standardize paroxysmal nocturnal hemoglobinuria (PNH) testing.
- Previous guidelines and studies informed the development and validation of a novel PNH assay.
Purpose of the Study:
- To develop and validate a robust, network-endorsed assay for PNH detection.
- To improve the precision and reliability of PNH testing across multiple laboratories.
Main Methods:
- A network analysis of over 20 samples was conducted to refine the assay protocol.
- Twenty-seven native samples underwent blind testing across participating laboratories, with central re-analysis.
- Assay validation included linearity, precision testing, and determination of assay limits (blank, detection, quantification).
Main Results:
- The final PNH assay protocol demonstrated comparable cell enumeration across laboratories, with no significant discrepancies.
- Biological sample characteristics influenced precision, necessitating adjustments like a red cell gating strategy to eliminate false positives.
- The validated assay achieved a Limit of Quantification of 0.01% for neutrophils and red cells, and 0.1% for monocytes.
Conclusions:
- A robust, network-validated PNH assay has been established, suitable for adoption by other laboratories.
- This assay facilitates high-resolution PNH analysis, enhancing diagnostic capabilities.
Related Concept Videos
Flow Cytometry
In...
Harmonic Mean
Take the example of the speed of a car, which is the measure of the rate of distance traveled. If the vehicle traverses the same distance back-and-forth, its average speed equals the total distance traveled divided by the total time taken. However, if the car moves with varying speeds, then the arithmetic mean is more skewed...