A biomechanical cadaver comparison of suture button fixation to plate fixation for pubic symphysis diastasis
Eric M Kiskaddon1, Amanda Wright2, Brett D Meeks1
1Orthopaedic Surgery, Wright State University Department of Orthopaedic and Plastic Surgery, Dayton, OH, United States.
Objectives:
To determine whether suture button fixation of the pubic symphysis is biomechanically similar to plate fixation in the treatment of partially stable pelvic ring injuries.
Methods:
Twelve pelvis specimens were harvested from fresh frozen cadavers. Dual-x-ray-absorptiometry (DXA) scans were obtained for all specimens. The pubic symphysis of each specimen was sectioned to simulate a partially stable pelvic ring injury. Six of the pelvises were instrumented using a 6 hole, 3.5 mm low profile pelvis plate and six of the pelvises were instrumented with two suture button devices. Biomechanical testing was performed on a pneumatic testing apparatus in a manner that simulates vertical stance. Displacement measurements of the superior, middle, and inferior pubic symphysis were obtained prior to loading, after an initial 440 N load, and after 30,000 and 60,000 rounds of cyclic loading. Statistical analysis was performed using Wilcoxon-Mann-Whitney tests, Fisher's exact test, and Cohen's d to calculate effect size. Significance was set at p < 0.05.
Results:
There was no difference between groups for DXA T scores (p = 0.749). Between group differences in clinical load to failure (p = 0.65) and ultimate load to failure (p = 0.52) were not statistically significant. For symphysis displacement, the change in fixation strength and displacement with progressive cyclic loading was not significant when comparing fixation types (superior: p = 0.174; middle: p = 0.382; inferior: p = 0.120).
Conclusion:
Suture button fixation of the pubic symphysis is biomechanically similar to plate fixation in the management of partially stable pelvic ring injuries.
Related Concept Videos
Fixation and Sectioning
The simplest type of preparation is the wet mount, in which the specimen is placed in a drop of liquid on the slide. A liquid specimen can be directly deposited on the slide using a dropper. Solid specimens, such as skin scraping, can be placed on the slide before adding a drop of liquid to prepare the wet mount. Sometimes the liquid is simply water, but stains are often added...
Carbon-dioxide Fixation
Sutures of the Skull
Sutures are immobile joints between adjacent bones of the skull. The narrow gap between the bones is filled with dense, fibrous connective tissue that unites the bones. The long sutures located between the skull bones are not straight but instead follow irregular, tightly twisting paths. These twisting lines tightly...
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Evolutionary Relationships through Genome Comparisons