Comparison of multi-institutional pre-treatment verification for VMAT of nasopharynx with delivery errors
E M Pogson1, S Arumugam2, C R Hansen3
1Institute of Medical Physics, School of Physics, Faculty of Science, The University of Sydney, Sydney, New South Wales, Australia; Liverpool and Macarthur Cancer Therapy Centres, Liverpool, NSW, Australia; Ingham Institute for Applied Medical Research, Sydney, NSW, Australia.
Purpose:
Measurement-based pre-treatment verification with phantoms frequently uses gamma analysis to assess acceptable delivery accuracy. This study evaluates the sensitivity of a commercial system to simulated machine errors for three different institutions' Volumetric Modulated Arc Therapy (VMAT) planning approaches.
Methods:
VMAT plans were generated for ten patients at three institutions using each institution's own protocol (manually-planned at institution 1; auto-planned at institutions 2 and 3). Errors in Multi-Leaf Collimator (MLC) field size (FS), MLC shift (S), and collimator angle (C) of -5, -2, -1, 1, 2 and 5 mm or degrees were introduced. Dose metric constraints discriminated which error magnitudes were considered unacceptable. The smallest magnitude error treatment plans deemed clinically unacceptable (typically for a 5% dose change) were delivered to the ArcCHECK for all institutions, and with a high-dose point ion chamber measurement in 2 institutions. Error detection for different gamma analysis criteria was compared.
Results:
Not all deliberately introduced VMAT plan errors were detected using a typical 3D 3%/3 mm global gamma pass rate of 95%. Considering all institutions, gamma analysis was least sensitive to negative FS errors. The most sensitive was a 2%/2 mm global analysis for institution 1, whilst for institution 2 it was 3%/3 mm global analysis. The majority of errors (58/59 for institution 1, 54/60 for institution 3) were detected using ArcCHECK and ion chamber measurements combined.
Conclusions:
Not all clinically unacceptable errors are detected. Combining ion chamber measurements with gamma analysis improved sensitivity and is recommended. Optimum gamma settings varied across institutions.
Related Concept Videos
Self-Evaluation: Self-Enhancement and Self-Verification
Strategies of Self-Presentation II: Self-Verification
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...
Fundamental Attribution Error
Random Error
Margin of Error