Purpose:
Intraoperative pelvic motion can alter the perceived cup inclination and version during non-navigated THA. We quantified pelvic motion during different phases of primary THA performed in the lateral decubitus through a posterolateral approach.
Methods:
Pelvic roll (rotation of the coronal plane) and pitch angles (rotation parallel to the coronal plane) were studied in 75 patients undergoing THA for osteoarthritis by four arthroplasty surgeons. Ten steps of surgery were defined. Angular motion was recorded with a miniature surgical device that utilizes inertial sensors.
Results:
The mean absolute roll ranged from 0.03° detected at the end of surgery to 4.13° detected during acetabular exposure. The mean absolute pitch ranged from 0.05° detected at the end of surgery to 2.54° detected during hip dislocation. The maximum pelvic roll and pitch detected during surgery averaged 17.62° (SD: 5.08) and 9.3° (SD: 3.39) respectively. Absolute roll and pitch angles were not affected by patient's BMI, sex, pre-operative hip motion, or surgeon. Before cup insertion, the greatest mean change in roll was observed during acetabular exposure (10.02° anteriorly), and for pitch was observed during dislocation (1.88° caudally).
Conclusion:
During THA performed through a posterolateral approach, there is a progressive anterior pelvic roll that peaks before cup insertion. This can lead to underestimation of cup anteversion during non-navigated THA. The anterior roll does not completely correct, even when all retractors and external forces acting on the pelvis are removed. Pelvic pitch that could affect the perceived cup inclination occurs to a lesser extent than pelvic roll.
Related Concept Videos
Errors in Global Positioning System
Sources of Self-Esteem II: Performance Feedback
Rolling Without Slipping
Rolling With Slipping
An object's rolling motion is characterized by its rotation around its axis, while linear motion refers to the object's translational motion along a surface. Frictional forces can...
Fundamental Attribution Error
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...