A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Single incision technique for implantation of subcutaneous implantable cardioverter defibrillators
Yousef H Darrat1, Francis Benn1, Mohsin Salih1
1Gill Heart Institute and VAMC, Cardiology, University of Kentucky, Lexington, KY, USA.
Background:
Subcutaneous implantable cardioverter defibrillators (S-ICDs) have gained increasing popularity because of certain advantages over transvenous ICDs. However, while conventional ICDs require a single surgical incision to implant, S-ICDS need two or three incisions, making them less appealing.
Objective:
This study sought out to investigate the feasibility of using a single-incision technique to implant S-ICDs.
Methods:
Patients qualifying for S-ICDs were considered for a single incision. A single incision is performed by making a left inframammary incision and then the subcutaneous tissue is dissected medially toward the lower sternum. Two sutures are placed in the fascia in the xiphoid area to anchor the lead and a tunneling tool is used to dissect the tissue to place the lead parallel to the sternum. Then subcutaneous tissues are dissected down the lateral chest wall over the muscle fascia to create the pulse generator pocket in the vicinity of the fifth and sixth intercostal spaces and near the mid-axillary line.
Results:
Eleven patients (six males and five females) successfully underwent S-ICD implantation with a single incision without acute complications (64% for primary prevention). The mean age is 47.4 ± 15.8 years. There were no lead dislodgements, inappropriate shocks, or any other issues during a median follow-up of 10 months (interquartile range 5-17). One patient had a successful appropriate shock for ventricular fibrillation about one year after device implant.
Conclusions:
A single incision for subcutaneous ICDs is feasible and safe in our early experience.
More Related Videos
13:35Endoscopic Bilateral Nipple-sparing Mastectomy via a Single Axillary Incision with Immediate Pre-pectoral Implant-based Breast Reconstruction
Published on: May 17, 2024
03:44A Unique Mouse Model for Quantitative Assessment of Biofilm Formation on Surgical Implants in Subcutaneous Abscess
Published on: June 6, 2025
Related Concept Videos
Load along a Single Axis
Consider a beam of length L subjected to a varying load, which is a combination of parabolic and trapezoidal load distribution along the x-axis. In this case, it is essential to determine the resultant loads, their locations, and...
Single Pipe Systems
In a Type I problem, fluid properties (density and viscosity), pipe characteristics (including diameter, length, and surface roughness), and the flow rate or average velocity are...
Angular Momentum: Single Particle
Single-pass Transmembrane Proteins
Inductance: Single-Phase And Three-Phase Line
Single-Phase Two-Wire Line:
A single-phase line consists of two solid cylindrical conductors, denoted as x and y. Each conductor carries phasor currents ix and iy, respectively. Given that the sum of these currents is...
Capacitance: Single-Phase And Three-Phase Line
Single-Phase Lines
Consider a single-phase, two-wire transmission line with equal phase spacing energized by a voltage source. One conductor carries a uniform positive charge, while the other carries an equal negative charge. The capacitance C of the line can be derived from the voltage V between the conductors. For a one-meter section of the line, the capacitance is given...