Humidification Performance of Passive and Active Humidification Devices Within a Spontaneously Breathing
Nobuto Nakanishi1, Jun Oto2, Taiga Itagaki1
1Emergency and Critical Care Medicine, Tokushima University Hospital, Tokushima, Japan.
Background:
Most heat-and-moisture exchangers (HMEs) for patients with tracheostomy and spontaneously breathing are small and have suction ports that allow some expiratory gas to escape, which loses water vapor held in the expired gas. Recently, a heated-and-humidified high-flow system for spontaneously breathing patients with tracheostomy was developed. Little is known, however, about the humidifying performance of HMEs or heated-and-humidified high-flow systems for spontaneous breathing patients with a tracheostomy.
Objective:
To investigate the humidifying performance of the HMEs and heated-and-humidified high-flow systems for spontaneously breathing patients with tracheostomy.
Methods:
Adult spontaneously breathing subjects with tracheostomy and were enrolled when their respiratory parameters and SpO were stable. We measured absolute humidity, relative humidity, and temperature by using a capacitance-type moisture sensor at the outlet of the tracheostomy tube. Heated-and-humidified high flow was delivered via the a humidifier and tracheostomy interface, and a selected HME. The subjects received heated-and-humidified high flow, after which an HME was used for humidification before switching back to a heated-and-humidified high-flow system.
Results:
Ten subjects (5 men, 5 women; mean ± SD age, 72 ± 12 y) were enrolled. The admission diagnoses were neurologic (5 subjects), respiratory failure (3), and cardiac arrest (2). The APACHE (Acute Physiology and Chronic Health Evaluation) II score was 24 (interquartile range, 20-27). Tracheostomy was performed on day 7 (interquartile range, 5-11 d) after endotracheal intubation, and the duration of mechanical ventilation was 10 d (interquartile range, 6-11 d). The temperature with the HME was 29.9 ± 1.0°C and, during heated-and-humidified high-flow use was 35.3 ± 0.8°C (P < .001). With both the HME and the heated-and-humidified high-flow system, the relative humidity reached 100%; the absolute humidity with HME was 30.2 ± 1.8 mg/L, and, with the heated-and-humidified high-flow system, was 40.3 ± 1.8 mg/L (P < .001).
Conclusions:
In spontaneously breathing subjects with tracheostomy, an heated-and-humidified high-flow system achieved higher absolute humidity than did an HME.
Related Concept Videos
Active versus Passive Immunity
Active Immunity
Active immunity refers to the resistance one develops...
Breathing
Spontaneity
Passive Filters
Low-Pass Filters
Low-pass filters are designed to transmit signals with frequencies lower than the cutoff frequency, ωc, and attenuate those above it. The cutoff...
Mechanism of Breathing I: Inspiration
The respiratory system, an essential network for breathing, comprises the conducting and respiratory zones, each playing a crucial role in the overall process of respiration. Let us explore the detailed mechanism of inspiration, or inhalation, which is the first phase of the respiratory cycle.
Pathway of Air during Inspiration
During inspiration, air enters our body through the nose or mouth and moves through the conducting zone,...
Mechanism of Breathing II: Expiration
Expiration, or exhaling, is a complex physiological process that begins as the inspiratory muscles begin to relax. This relaxation triggers a series of events that epitomize the efficiency of the respiratory system.
Mechanism of Expiration: