Prospective Comparative Study in Spine Surgery Between O-Arm and Airo Systems: Efficacy and Radiation Exposure
Kaissar Farah1, Pierre Coudert2, Thomas Graillon1
1Department of Neurosurgery, La Timone University Hospital, Marseille, France; Spine Unit, La Timone University Hospital, Marseille, France.
Objective:
Pedicle screw placement remains challenging. The present study focuses on the comparison between 2 intraoperative-based neuronavigation systems (O-Arm and AIRO) during thoracolumbar screw instrumentation.
Methods:
This is a prospective, comparative, nonrandomized study conducted in 2 French academic centers. The O-Arm was used at the University Hospital of Bordeaux, whereas the AIRO was used at the University Hospital of Marseille. Routine computed tomography was performed on postoperative day 2 to evaluate pedicle screw placement. Measures of radiation exposure were extracted directly from reports provided by each system. The effective dose was calculated.
Results:
Overall, 74 screws were placed in 11 patients in the O-Arm group and 84 in 11 patients in the AIRO group. In the first group, 90.8% were rated as acceptable and 92.2% in the second (P > 0.05) according to the Heary and Gertzbein classifications, respectively, for thoracic and lumbar spine. Differences between both implantation systems were significant (P < 0.05) concerning dose length product (235 and 1039 mGy/cm, in O-Arm and AIRO, respectively), overall mean radiation dose received by 1 patient (3.52 and 15.6 mSv in O-Arm and AIRO, respectively), mean radiation dose per single scan (2.58 and 8.7 mSv in O-Arm and AIRO, respectively), mean effective dose per instrumented level (1.04 and 3.9 mSv in O-Arm and AIRO, respectively), and radiation dose received by the primary surgeon (0.63 and 0 μSv in O-Arm and AIRO, respectively).
Conclusions:
Intraoperative computed tomography-based navigation is a major innovation that improves the accuracy of pedicle screw positioning with acceptable patient radiation exposure and reduced surgical team exposure.
More Related Videos
08:34Optimizing Minimally Invasive Spine Surgery: A Fully 3D CT O-Arm Navigated Workflow in MIS TLIF
Published on: October 17, 2025
03:07Single-Port Robotic-assisted Transaxillary Breast-conserving Surgery: A Prospective, Single-arm, Non-randomized Phase IIa Clinical Trial
Published on: August 19, 2025
Related Concept Videos
Comparative Excretory Systems
Biological Effects of Radiation
Muscles that Move the Arm
The pectoralis major has two origins. Its clavicular head originates on the medial half of the clavicle. In contrast, the sternocostal head originates on the costal cartilages of ribs 1-6, the sternum, and the aponeurosis of the external oblique of the...
Self-Efficacy
Radiation: Applications
The average...
Absorption of Radiation