Extent of tube-current reduction using sinogram affirmed iterative reconstruction in pediatric computed tomography:
1Department of Radiologic Sciences, Kuwait University, PO Box 31470, Sulaibikhat, 90805, Kuwait. ajit@hsc.edu.kw.
Background:
Iterative image reconstruction techniques can produce diagnostic-quality computed tomography (CT) images with lower radiation dose.
Objective:
To quantify the reduction in x-ray tube-current setting and optimize pediatric CT scans using different strengths of an iterative reconstruction technique.
Materials And Methods:
The head, chest and abdomen regions of an anthropomorphic phantom representing a 5-year-old patient were scanned using standard CT protocols. Images were reconstructed using filtered back projection and different strengths of a sinogram affirmed iterative reconstruction technique. Repeated measurements of contrast-to-noise ratios in the lungs, bone and soft-tissue regions of the phantom were carried out. Maximum increase in contrast-to-noise ratio with iterative reconstruction strength was identified and a tube-current reduction factor was calculated. Head scans were repeated with reduced tube current and compared to filtered back projection images.
Results:
Iterative reconstruction strength of 3 for head and chest images and 4 for abdomen images were optimum, resulting in contrast-to-noise ratio increase of about 50%. A tube-current reduction factor of 1.2 for head images was calculated. Images of the head acquired using reduced tube-current showed similar contrast-to-noise ratio as images form filtered back projection with full tube current.
Conclusion:
Optimum strength of iterative reconstruction technique has been identified for head, chest and abdomen images. Reductions in tube current of 20%, resulting in similar radiation dose reduction, have been established.
More Related Videos
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...
Oxidation-Reduction Reactions
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Uterine Tubes