Performance evaluation of serum IgG subclass quantification using a SPAPLUS turbidimetric analyzer and comparison
Eun Hye Cho1, Rihwa Choi2, Eun-Suk Kang1
1a Department of Laboratory Medicine & Genetics , Samsung Medical Center, Sungkyunkwan University School of Medicine , Seoul , Korea.
Abstract:
IgG consists of four subclasses: IgG1, IgG2, IgG3, and IgG4. Changes in the serum concentration of each subclass reflect different clinical situations, and quantification of each subclass is important to assess patients' clinical states. Herein, we evaluated the analytical performance of the SPAPLUS turbidimetric analyzer (The Binding Site, Birmingham, UK) for IgG subclass. Precision, linearity, comparison with the BNII system (Siemens Healthineers, Erlangen, Germany), and reference interval were assessed according to the Clinical and Laboratory Standards Institute (CLSI) guidelines. The repeatability and within-laboratory precision were within 5% for all IgG subclasses. The coefficient of determination (R2) was higher than 0.99 for the analytical measurement range in all IgG subclasses. Comparison between SPAPLUS and BNII revealed significant differences in IgG1, IgG3, and IgG4 (p < .0001). IgG1 and IgG4 values were lower in SPAPLUS than BNII. On the other hand, IgG3 values were higher in SPAPLUS than BNII. The SPAPLUS turbidimetric analyzer exhibited good analytical performance for quantification of four IgG subclasses. Because of the differences between SPAPLUS and BNII, follow-up test for disease monitoring should be performed with same instrument.
More Related Videos
Related Concept Videos
Transcytosis of IgG
IgG molecules from a mother undergo transcytosis starting around 13 weeks of gestation. The amount of IgG transferred and entering the fetal blood circulation increases with...
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Evolutionary Relationships through Genome Comparisons
Comparison Between Electrical And Gravitational Forces
Since both are inverse square law forces, the distance gets canceled when the ratio of the two forces is considered. Instead, the ratio of the electrical and gravitational forces depends on...
Comparison between RL and RC circuits