Purpose:
Elimination of pace-capture along pulmonary vein isolation (PVI) lesion sets reduces atrial fibrillation (AF) recurrence in catheter ablation of paroxysmal AF. Pacing from the RF ablation electrode during RF application is prevented within the CARTO electroanatomic mapping system (Biosense Webster, Inc.) due to theoretical safety considerations. We evaluated a method of pacing the distal ablation electrode during RF application in the CARTO system, thus avoiding repeated activation and inactivation of the pacing channel and facilitating immediate recognition of pace-capture loss. We investigated the safety, feasibility, and utility of simultaneous pace-ablate (SPA) during AF ablation with the CARTO-3 system and a contact-force sensing RF ablation catheter.
Methods:
Safety of feasibility of SPA was evaluated in 250 patients undergoing first-time AF ablation. Frequency and regional distribution of pace-capture following PVI was evaluated in a cohort of 50 consecutive patients undergoing catheter ablation of paroxysmal AF.
Results:
SPA was successfully performed in all 250 patients without adverse event. At least one pace-capture site was noted in 22 of 50 PAF patients (44%), and pace-capture following PVI was most common at anterior and superior left atrial sites. There were 2.0 ± 3.3 RF applications during pacing via the distal ablation electrode per patient, and all lesions sets were successfully rendered unexcitable.
Conclusions:
Pace-capture along the completed PVI lesion set remains common despite utilization of contact-force sensing RF ablation catheters and automated lesion annotation. Simultaneous pace-ablate in AF ablation using the CARTO system may be safely used to render atrial lesion sets unexcitable.
Related Concept Videos
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Contact Angle
The adhesive force is the molecular force between molecules of different materials, that is, between the molecules of the solid and the liquid. The cohesive...
Contact-dependent Signaling
Gap Junctions
In animal cells, gap junctions are formed...
Electromotive Force
Introduction to Special Senses
Veins
Structure of Veins:
The structure of veins is specifically designed to assist in the low-pressure transportation of...