A Quantitative Assessment of The Yeast Lipidome using Electrospray Ionization Mass Spectrometry
Published on: August 21, 2009
Optimization of Electrospray Ionization Source Parameters for Lipidomics To Reduce Misannotation of In-Source
Rose M Gathungu1, Pablo Larrea1, Matthew J Sniatynski1
1Department of Medicine , Division of Sleep and Circadian Disorders, Brigham and Women's Hospital, and Department of Medicine, Division of Sleep Medicine, Harvard Medical School , Boston , Massachusetts 02115 , United States.
Abstract:
Lipidomics requires the accurate annotation of lipids in complex samples to enable determination of their biological relevance. We demonstrate that unintentional in-source fragmentation (ISF, common in lipidomics) generates ions that have identical masses to other lipids. Lysophosphatidylcholines (LPC), for example, generate in-source fragments with the same mass as free fatty acids and lysophosphatidylethanolamines (LPE). The misannotation of in-source fragments as true lipids is particularly insidious in complex matrixes since most masses are initially unannotated and comprehensive lipid standards are unavailable. Indeed, we show such LPE/LPC misannotations are incorporated in the data submitted to the National Institute of Standards and Technology (NIST) interlaboratory comparison exercise. Computer simulations exhaustively identified potential misannotations. The selection of in-source fragments of highly abundant lipids as features, instead of the correct recognition of trace lipids, can potentially lead to (i) missing the biologically relevant lipids (i.e., a false negative) and/or (ii) incorrect assignation of a phenotype to an incorrect lipid (i.e., false positive). When ISF is not eliminated in the negative ion mode, ∼40% of the 100 most abundant masses corresponding to unique phospholipids measured in plasma were artifacts from ISF. We show that chromatographic separation and ion intensity considerations assist in distinguishing precursor ions from in-source fragments, suggesting ISF may be especially problematic when complex samples are analyzed via shotgun lipidomics. We also conduct a systematic evaluation of electrospray ionization (ESI) source parameters on an Exactive equipped with a heated electrospray ionization (HESI-II) source with the objective of obtaining uniformly appropriate source conditions for a wide range of lipids, while, at the same time, reducing in-source fragmentation.
Related Concept Videos
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Sinusoidal Sources
In homes, the power supplies use sinusoidal sources to provide electricity. These sources generate a voltage that varies sinusoidally...
AC Sources
Sources of Law
Constitutional law is foundational, deriving from federal and state constitutions, and...
Source Transformation
It is essential to note that when...
Multiple Voltage Sources
In series, the positive terminal of one battery is connected to the negative terminal of another battery. Hence, the voltage of each battery is added to give the net voltage, which is increased because each battery boosts the electrons that enter it. The same current flows through each battery because they are connected in series.
Batteries are...