Related Experiment Video
Updated: Feb 4, 2026

Ferromagnetic Bare Metal Stent for Endothelial Cell Capture and Retention
Published on: September 18, 2015
Experimental Direct Measurement of Clot-Capturing Ability of Stent Retrievers
Tomotaka Ohshima1, Reo Kawaguchi2, Yoshitaka Nagano3
1Neuroendovascular Therapy Center, Aichi Medical University, Nagakute, Aichi, Japan.
Background:
Stent retrievers (SRs) can be used to perform mechanical thrombectomy for the treatment of acute major arterial occlusion. Recanalization is faster, and outcomes are better with treatment involving these devices than with internal treatment. Although several SRs are available, their clot-capturing abilities are unclear. Therefore in the present study, we numerically evaluated the clot-capturing abilities of SRs in an experimental vascular model.
Methods:
A sham clot (urethane foam) was fixed with sutures at the middle of a vascular model (polyvinyl chloride tube). One end of the tube was connected to a measuring instrument. From the other end, an SR was inserted and deployed over the sham clot. The delivery wire of the stent was then withdrawn at a constant velocity using an automatic withdrawal machine. The maximum frictional force before the stent left the clot was measured. Five stents (Trevo ProVue [2 sizes], Revive, and Solitaire [2 sizes]) and three stent-deployment techniques (standard, push-and-fluff, and wire-push techniques) were evaluated.
Results:
The clot-capturing ability (maximum withdrawing force [N: newton]) was greater for large-diameter stents than for small-diameter stents (mean 0.39 ± 0.11 vs. 0.56 ± 0.18). For Trevo and Revive, the clot-capturing ability was highest with the push and fluff technique (mean 0.43 ± 0.05). For Solitaire, the ability was the highest with the simple wire-push technique (mean 0.705 ± 0.16).
Conclusions:
We successfully numerically evaluated the clot-capturing abilities of SRs. The clot-capturing ability differed among SRs and among stent-deployment techniques.
Related Concept Videos
Retrieval
Recall involves accessing information without cues, such as during an essay test, where individuals must retrieve facts and concepts from memory unaided. Another example is remembering the name of a colleague...
ER Retrieval Pathway
The ER uses many checkpoints to prevent the entry of incorrectly folded or a resident protein as cargo onto a transport vesicle. These mechanisms...
Clot Retraction and Fibrinolysis
Microbial Growth Measurement: Direct Methods
Directing Effect of Substituents: meta-Directing Groups
Experimental Determination of Chemical Formula