CT-Guided Percutaneous Spine Biopsy Specimen Adequacy, Pathology Concordance, and Negative Predictive Value with
Jeffrey P Guenette1, Charles H Cho1, Raymond Y Huang1
1Department of Radiology, Brigham and Women's Hospital, Harvard Medical School, Boston, MA.
Objective:
Determine computed tomography-guided percutaneous spine biopsy specimen adequacy, pathology-imaging concordance, and negative predictive value with battery-powered drill vs manual approach.
Materials And Methods:
One-hundred-fourteen consecutive computed tomography-guided percutaneous spine biopsies in 109 patients (age: 61.1 ± 15.4 years; range: 17-90 years; males: 55, 50.5%; females: 54, 49.5%) performed at a single institution from September 2013 through January 2017 were retrospectively reviewed. Specimen adequacy was recorded. Imaging-pathology concordance was assessed. Chi-square tests compared specimen adequacy and imaging-pathology concordance obtained with a battery-powered drill vs manual approach. Negative predictive values were calculated.
Results:
Battery-powered drill yielded slightly better, but not statistically significant, specimen adequacy (96% vs 90% overall, P = 0.270; 96% vs 89% for suspected neoplasm, P = 0.278; 95% vs 90% for suspected infection, P = 0.514), pathology-imaging concordance (82% vs 74% overall, P = 0.301; 92% vs 77% for suspected neoplasm, P = 0.107; 71% vs 65% for suspected infection, P = 0.602), and negative predictive value (65% vs 41% overall; 75% vs 33% for suspected neoplasm; 58% vs 33% for suspected infection). Four battery-powered drill procedures were technically unsuccessful.
Conclusions:
Use of a battery-powered drill appears to yield similar to slightly better spine biopsy specimens than a manual approach, but also appears to carry a greater risk of technical failure. The battery-powered drill may be particularly helpful for procedures with complex approaches, but trajectory planning remains of paramount importance.
More Related Videos
08:34Optimizing Minimally Invasive Spine Surgery: A Fully 3D CT O-Arm Navigated Workflow in MIS TLIF
Published on: October 17, 2025
11:27Preparation and Respirometric Assessment of Mitochondria Isolated from Skeletal Muscle Tissue Obtained by Percutaneous Needle Biopsy
Published on: February 7, 2015
Related Concept Videos
Batteries and Fuel Cells
Kendall's Coefficient of Concordance
DC Battery
Predicting Molecular Geometry
Power
Negative Regulator Molecules