Dosimetry for Cell Irradiation using Orthovoltage 40-300 kV X-Ray Facilities
Published on: February 20, 2021
Monte Carlo and 60 Co-based kilovoltage x-ray dosimetry methods
Michael J Lawless1, Lianna Dimaso2, Benjamin Palmer2
1Department of Human Oncology, University of Wisconsin-Madison, Madison, WI, 53705, USA.
Purpose:
This work seeks to investigate new methods to determine the absorbed dose to water from kilovoltage x rays. Current methods are based on measurements in air and rely on correction factors in order to account for differences between the photon spectrum in air and at depth in phantom, between the photon spectra of the calibration beam and the beam of interest, or in the radiation absorption properties of air and water. This work aims to determine the absorbed dose to water in the NIST-matched x-ray beams at the University of Wisconsin Accredited Dosimetry Calibration Laboratory (UWADCL). This will facilitate the use of detectors in terms of dose to water, which will allow for a simpler determination of dose to water in clinical kilovoltage x-ray beams.
Materials And Methods:
A model of the moderately filtered x-ray beams at the UWADCL was created using the BEAMnrc user code of the EGSnrc Monte Carlo code system. This model was validated against measurements and the dose to water per unit air kerma was calculated in a custom built water tank. Using this value and the highly precise measurement of the air kerma made by the UWADCL, the dose to water was determined in the water tank for the x-ray beams of interest. The dose to water was also determined using the formalism defined in the report of AAPM Task Group 61 and using a method that makes use of a Co absorbed dose-to-water calibration coefficient and a beam quality correction factor to account for differences in beam quality between the Co calibration and kilovoltage x-ray beam of interest. The dose to water values as determined by these different methods was then compared.
Results:
The BEAMnrc models used in this work produced simulations of transverse and depth dose profiles that agreed with measurements with a 2%/2 mm criteria gamma test. The dose to water as determined from the different methods used here agreed within 3.5% at the surface of the water tank and agreed within 1.8% at a depth of 2 cm in phantom. The dose-to-water values all agreed within the associated uncertainties of the methods used in this work. Both the Monte Carlo-based method and the Co-based method had a lower uncertainty than the TG-61 methodology for all of the x-ray beams used in this work.
Conclusion:
Two new dose determination methods were used to determine the dose to water in the NIST-matched x-ray beams at the UWADCL and they showed good agreement with previously established techniques. Due to the improved Monte Carlo calculation techniques used in this work, both of the methods have lower uncertainties compared to TG-61. The methods presented in this work compare favorably with calorimetry-based standards established at other institutions.
Related Concept Videos
X-ray Crystallography
Diffraction
Diffraction is the change in the direction of travel experienced by an electromagnetic wave when it encounters a physical barrier whose dimensions are comparable to those of the wavelength of the light. X-rays are electromagnetic radiation with wavelengths about as long as the distance between neighboring...
X-ray Imaging
X-ray Diffraction of Biological Samples
According to Bragg's law, when X-rays strike the sample positioned on a stage, the rays are scattered by the electron clouds around the sample atoms. The X-ray diffraction or scattering is caused by constructive interference of the X-ray waves that reflect off the internal...
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Lewis Acids and Bases
A coordinate covalent bond (or dative bond) occurs when one of the atoms in the bond provides both bonding electrons. For example, a coordinate covalent bond occurs when a water molecule combines with a hydrogen ion to form a hydronium ion. A coordinate covalent bond also results when...