Online Gas-Free Electrodialytic KOH Eluent Generator for Ion Chromatography
Yifei Lu1, Liting Zhou1, Bingcheng Yang1
1Shanghai Key Laboratory of New Drug Design, School of Pharmacy , East-China University of Science and Technology , Shanghai 200237 , China.
Abstract:
An online gas-free KOH electrodialytic eluent generator (EDG) with two-membrane configuration is described for ion chromatography (IC). A central eluent channel is separated from two outer regenerant channels bearing KOH solution (or one is water) by stacked cation-exchange membranes (sCEMs) and a bipolar membrane (BPM) plus stacked CEMs (BPM-sCEMs), in which the anion-exchange membrane (AEM) of BPM is facing the central channel (the anode direction). Independent fluid input or output ports address all channels. One platinum screen electrode is put in each outer channel, and the sCEM side is anode with respect to the cathode of the BPM-sCEM side. Under the electric field, enhanced water splitting at the intermediate layer of BPM will occur, offering hydroxide and hydronium; hydroxide electromigrates through AEM into the central channel, while the potassium of KOH feed solution at the anode migrates into the central channel to combine with hydroxide to form a solution of KOH. Because the central eluent channel is spatially isolated from both electrodes, the generated KOH solution is gas-free and no gas-removal device is required. More important, the AEM side of BPM is contactless with the alkaline solution, nearly avoiding possible AEM degradation when immersed into concentrated alkaline solution. This ensures long-term running stability of the EDG and high purity of the produced KOH solution (as indicated by the typical suppressed background conductance of 0.28 μS/cm). The EDG has a pressure tolerance (at least 21 MPa), and the produced KOH concentration is up to 101 μeq/min with near-unity faradaic efficiency. When operated in both isocratic and gradient modes, the EDG demonstrates excellent reproducibility, as indicated by the retention time of RSD ≤ 0.08% and the peak area of RSD ≤ 0.6%. To the best of our knowledge, this is the first description of gas-free EDG matched conventional IC system.
More Related Videos
08:22Nitrogen Compound Characterization in Fuels by Multidimensional Gas Chromatography
Published on: May 15, 2020
11:44Qualitative Characterization of the Aqueous Fraction from Hydrothermal Liquefaction of Algae Using 2D Gas Chromatography with Time-of-flight Mass Spectrometry
Published on: March 6, 2016
Related Concept Videos
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...
Ion-Exchange Chromatography
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...