Chromium and Nickel Concentrations in Subjects with a Stainless Steel Metal-on-Metal Cervical Disc Arthroplasty:
Vaneet Singh1, Anastasia K Skipor2, Abdulhafez A Selim1
1Medtronic, Inc., Memphis, Tennessee.
Background:
Cervical disc arthroplasty (CDA) has emerged as an alternative to anterior cervical discectomy and fusion for degenerative cervical disc disease. The artificial discs provide intervertebral motion using multicomponent articulation and thus tend to generate particulate debris and soluble metal ions. Limited information is available on the long-term metal concentrations and associated systemic adverse events observed in metal-on-metal CDA. Serum chromium (Cr) and nickel (Ni) concentrations were assessed in patients implanted with ball-in-trough stainless steel-based cervical disc through 7 years.
Methods:
A prospective, nonrandomized longitudinal study was conducted that included 25 patients following rigorous exclusion criteria that included no previous permanent metal implants and no professional exposure to metal particles. Blood serum Cr and Ni concentrations were assayed preoperatively and at 3, 6, 12, 24, 36, 60, and 84 months postoperatively using high-resolution inductively coupled plasma-mass spectrometry. Longitudinal statistical comparisons were made using the Friedman test with statistical significance at P < .05.
Results:
Median serum concentrations determined preoperatively and at 3, 6, 12, 24, 36, 60, and 84 months postoperatively were 0.074, 0.106, 0.132, 0.170, 0.172, 0.274, 0.192, and 0.203 ng/mL for Cr and 0.085, 0.178, 0.222, 0.175, 0.205, 0.284, 0.181, and 0.194 ng/mL for Ni. The serum Cr concentrations were statistically higher for all postoperative time periods compared to preoperative concentration (Friedman P <.01), whereas serum Ni concentration was statistically higher at the 84-month postoperative time period than the preoperative concentration (Friedman P <.01) and then the concentration at 3, 12, 24, and 60 months postoperatively (Friedman P < .03).
Conclusions:
The Cr concentrations detected at all postoperative times were statistically higher than preoperative concentrations, whereas Ni concentration was statistically higher than the preoperative concentration only at 84 months.
Related Concept Videos
Properties of Transition Metals
Bonding in Metals
Metallic Solids
All metallic solids exhibit high thermal and electrical conductivity, metallic luster, and malleability....
Alkali Metals
Table 1: Properties of the alkali metals
Metal-Ligand Bonds
In these complexes, transition metals form coordinate covalent bonds, a kind of Lewis acid-base interaction in which both of the electrons in the bond are contributed by a donor (Lewis base) to an electron acceptor (Lewis acid). The Lewis acid in...
Theory of Metallic Conduction
In this theory, Newton's second law of motion is used to determine the acceleration of an electron in the presence of an applied electric field. Then, its velocity is expressed via this acceleration.
An electron moves through the crystal, containing positive ions,...